Korean J Clin Microbiol.  2008 Apr;11(1):18-22. 10.5145/KJCM.2008.11.1.18.

Evaluation of MicroScan MICroSTREP Plus Antimicrobial Susceptibility Panel for Testing Streptococcus pneumoniae

Affiliations
  • 1Department of Laboratory Medicine, Hallym University College of Medicine, Seoul, Korea. jaeseok@hallym.or.kr
  • 2Department of Laboratory Medicine, Hallym University Sacred Heart Hospital, Anyang, Korea.

Abstract

BACKGROUND: The MicroScan MICroSTREP plus panel for susceptibility testing of various streptococci, including Streptococcus pneumoniae, has recently been introduced in Korea. The current study evaluated the usefulness of MicroScan MICroSTREP plus panel for antimicrobial susceptibility test of S. pneumoniae.
METHODS
A total of 75 clinical isolates of S. pneumoniae were tested for antimicrobial susceptibility to penicillin, cefotaxime, ceftriaxone, meropenem, vancomycin, clindamycin, erythromycin, and levofloxacin with the MicroScan MICroSTREP plus panel and clinical and laboratory standard institute (CLSI) reference broth microdilution method. For 46 of 75 isolates, additional susceptibility tests to penicillin and cefotaxime were performed with Etest.
RESULTS
The overall essential agreement of MICs (within one dilution of MICs) defined by the MicroScan MICroSTREP plus panel and reference method was 93.0%. Overall there were 11.7% minor, 0.7% major, and 0.7% very major interpretative category errors observed. The results of antibiotic susceptibility testing by Etest were similar to those obtained by the MicroScan MICroSTREP plus panel.
CONCLUSION
The MicroScan MICroSTREP plus panel, a commercial broth microdilution method, has a comparable accuracy to CLSI broth microdilution method for the resistance testing of S. pneumonia. This panel can be used for determining susceptibilities of S. pneumoniae to a wide variety of antimicrobial agents in clinical microbiology laboratories.

Keyword

Streptococcus pneumoniae; Antimicrobial susceptibility test; MicroScan; MICroSTREP plus panel

MeSH Terms

Anti-Infective Agents
Cefotaxime
Ceftriaxone
Clindamycin
Erythromycin
Korea
Ofloxacin
Penicillins
Pneumonia
Streptococcus
Streptococcus pneumoniae
Thienamycins
Vancomycin
Anti-Infective Agents
Cefotaxime
Ceftriaxone
Clindamycin
Erythromycin
Ofloxacin
Penicillins
Thienamycins
Vancomycin

Reference

References

1. Spellerberg B and Brandt C. Streptococcus. Murray PR, Baron EJ, Jorensen JH, Landry ML, Pfaller MA, editors. Manual of clinical microbiology. 9th ed.Washington, D.C.: ASM Press;2006. p. 412–9.
2. Jorgensen JH, McElmeel ML, Crawford SA. Evaluation of the Dade MicroScan MICroSTREP antimicrobial susceptibility testing panel with selected Streptococcus pneumoniae challenge strains and recent clinical isolates. J Clin Microbiol. 1998; 36:788–91.
3. Beekmann SE, Heilmann KP, Richter SS, Garcia-de-Lomas J, Doern GV. the GRASP study group. Antimicrobial resistance in Streptococcus pneumoniae, Haemophilus influenzae, Moraxella catarrhalis and group A B-hemolytic streptococci in 2002–03; Results of the multinational GRASP surveillance program. Int J Antimicrob Agents. 2005; 25:148–156.
4. Doern GV, Richter SS, Miller A, Miller N, Rice C, Heilmann K, et al. Antimicrobial resistance among Streptococcus pneumoniae in the United States: have we begun to turn the corner on resistance to certain antimicrobial classes? Clin Infect Dis. 2005; 41:139–48.
5. Guthrie LL, Banks S, Setiawan W, Waites KB. Comparison of MicroScan MICroSTREP, PASCO, and Sensititre MIC panels for determining antimicrobial susceptibilities of Streptococcus pneumoniae. Diagn Microbiol Infect Dis. 1999; 33:267–73.
Article
6. Clinical and Laboratory Standards Institute. Performance standards for antimicrobial susceptibility testing; sixteenth informational supplement. CLSI document M100-S16. Wayne, PA; CLSI. 2006.
7. Clinical and Laboratory Standards Institute. Methods for dilution antimicrobial susceptibility tests for bacteria that grow aerobically; approved standard–seventh edition. CLSI document M7-A7. Wayne, PA; CLSI. 2006.
8. Jorgensen JH, Ferraro MJ, McElmeel ML, Spargo J, Swenson JM, Tenover FC. Detection of penicillin and extended-spectrum cephalosporin resistance among Streptococcus pneumoniae clinical isolates by use of the E test. J Clin Microbiol. 1994; 32:159–63.
Article
9. Huh JW, Nah J, Lee SH, Pai CH. Evaluation of the E test for penicillin and cefotaxime susceptibility testing of Streptococcus pneumoniae. Korean J Lab Med. 1996; 16:330–9.
10. Diekema DJ, Beach ML, Pfaller MA, Jones RN. SENTRY Participants Group. Antimicrobial resistance in viridans group streptococci among patients with and without the diagnosis of cancer in the USA, Canada and Latin America. Clin Microbiol Infect. 2001; 7:152–7.
Article
11. Lee H, Kim CK, Lee J, Lee SH, Ahn JY, Hong SG, et al. Antimicrobial resistance of clinically important bacteria isolated from 12 hospitals in Korea in 2005 and 2006. Korean J Clin Microbiol. 2007; 10:59–69.
Full Text Links
  • KJCM
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles