J Korean Ophthalmol Soc.  2017 Mar;58(3):276-282. 10.3341/jkos.2017.58.3.276.

Comparison of Central Corneal Thickness Measured by Swept-source Optical Coherence Tomography and Ultrasound Pachymetry

Affiliations
  • 1Siloam Eye Hospital, Seoul, Korea. jhkim@siloam.co.kr
  • 2The Institute of Vision Research, Department of Ophthalmology, Yonsei University College of Medicine, Seoul, Korea.

Abstract

PURPOSE
To analyze and compare ultrasound pachymetry (USP) with a more recently adopted device, the intra ocular lens (IOL) master 700, which are both used to measure central corneal thickness.
METHODS
The central corneal thickness was measured in 24 eyes of 12 glaucoma patients and in 83 eyes of 42 normal patients. First, the IOL master 700 was used to measure the central corneal thickness, followed by measurements taken using USP later. The results were analyzed using a paired t-test. We analyzed the agreement and the correlations between the two test devices by using Bland-Altman plots and the Pearson correlation test. To evaluate the reproducibility, measurements with the IOL master 700 were taken twice for a few normal patients within a small time interval.
RESULTS
Via the IOL master 700, the thickness of the central cornea showed a high reproducibility and repeatability, demonstrating 2.7 ± 1.7 µm for the test-retest variability, 6.78% for the coefficient of variation, and 0.997 for the intraclass correlation value. The mean measurements using USP and the IOL master 700 are 554.4 ± 37.4 µm and 551.1 ± 37.1 µm, respectively, showing that the IOL master 700 measured significantly smaller values than USP with a p-value < 0.001. The deviations between the two methods are scattered throughout the 95% confidence interval. According to the Pearson correlation test, the measured values of the two test devices were found to have a highly positive correlation (r = 0.977, p < 0.0001).
CONCLUSIONS
This study demonstrated that the central corneal thickness (CCT) measured via the IOL master was significantly thinner than that of USP, and the two test devices had a high correlation and good agreement. The CCT value measured via the IOL master 700 also exhibited high reproducibility.

Keyword

Central corneal thickness; Ultrasound pachymetry; IOL master 700

MeSH Terms

Cornea
Glaucoma
Humans
Tomography, Optical Coherence*
Ultrasonography*

Figure

  • Figure 1 Bland Altman plots between the 2 methods. The middleline is the mean and the lines on the side represent the upper and lower 95% limits of agreement. USP = ultrasound pachymetry; IOL = intraocular lens; SD = standard deviation.

  • Figure 2 Correlation analysis graph (Scatter plot) between the 2 methods. Scattergram showing the correlation of central corneal thickness measured by USP and IOL master 700. USP = ultrasound pachymetry; IOL = intraocular lens.


Reference

1. Chihara E. Assessment of true intraocular pressure: the gap between theory and practical data. Surv Ophthalmol. 2008; 53:203–218.
2. Ou RJ, Shaw EL, Glasgow BJ. Keratectasia after laser in situ keratomileusis (LASIK): evaluation of the calculated residual stromal bed thickness. Am J Ophthalmol. 2002; 134:771–773.
3. Wang Z, Chen J, Yang B. Posterior corneal surface topographic changes after laser in situ keratomileusis are related to residual corneal bed thickness. Ophthalmology. 1999; 106:406–409. discussion 409-10.
4. Jonas JB, Stroux A, Velten I, et al. Central corneal thickness correlated with glaucoma damage and rate of progression. Invest Ophthalmol Vis Sci. 2005; 46:1269–1274.
5. Grulkowski I, Liu JJ, Zhang JY, et al. Reproducibility of a long-range swept-source optical coherence tomography ocular biometry system and comparison with clinical biometers. Ophthalmology. 2013; 120:2184–2190.
6. Harper CL, Boulton ME, Bennett D, et al. Diurnal variations in human corneal thickness. Br J Ophthalmol. 1996; 80:1068–1072.
7. Copt RP, Thomas R, Mermoud A. Corneal thickness in ocular hypertension, primary open-angle glaucoma, and normal tension glaucoma. Arch Ophthalmol. 1999; 117:14–16.
8. Ling T, Ho A, Holden BA. Method of evaluating ultrasonic pachometers. Am J Optom Physiol Opt. 1986; 63:462–466.
9. Li EY, Mohamed S, Leung CK, et al. Agreement among 3 methods to measure corneal thickness: ultrasound pachymetry, Orbscan II, and Visante anterior segment optical coherence tomography. Ophthalmology. 2007; 114:1842–1847.
10. Ventura AC, Wälti R, Böhnke M. Corneal thickness and endothelial density before and after cataract surgery. Br J Ophthalmol. 2001; 85:18–20.
11. Savini G, Carbonelli M, Barboni P, Hoffer KJ. Repeatability of automatic measurements performed by a dual Scheimpflug analyzer in unoperated and post-refractive surgery eyes. J Cataract Refract Surg. 2011; 37:302–309.
12. Olsen T, Thorwest M. Calibration of axial length measurements with the Zeiss IOLMaster. J Cataract Refract Surg. 2005; 31:1345–1350.
13. Choi JH, Roh GH. The reproducibility and accuracy of biometry parameter measurement from IOL Master (R). J Korean Ophthalmol Soc. 2004; 45:1665–1673.
14. Shin JA, Chung SK. Comparison of the refractive results measured by ultrasound and partial coherence interferometers. J Korean Ophthalmol Soc. 2013; 54:723–727.
15. Maldonado MJ, Ruiz-Oblitas L, Munuera JM, et al. Optical coherence tomography evaluation of the corneal cap and stromal bed features after laser in situ keratomileusis for high myopia and astigmatism. Ophthalmology. 2000; 107:81–87. discussion 88.
16. Muscat S, McKay N, Parks S, et al. Repeatability and reproducibility of corneal thickness measurements by optical coherence tomography. Invest Ophthalmol Vis Sci. 2002; 43:1791–1795.
17. Izatt JA, Hee MR, Swanson EA, et al. Micrometer-scale resolution imaging of the anterior eye in vivo with optical coherence tomography. Arch Ophthalmol. 1994; 112:1584–1589.
18. Radhakrishnan S, Rollins AM, Roth JE, et al. Real-time optical coherence tomography of the anterior segment at 1310 nm. Arch Ophthalmol. 2001; 119:1179–1185.
19. Bovelle R, Kaufman SC, Thompson HW, Hamano H. Corneal thickness measurements with the Topcon SP-2000P specular microscope and an ultrasound pachymeter. Arch Ophthalmol. 1999; 117:868–870.
20. Módis L Jr, Langenbucher A, Seitz B. Corneal thickness measurements with contact and noncontact specular microscopic and ultrasonic pachymetry. Am J Ophthalmol. 2001; 132:517–521.
21. Yaylali V, Kaufman SC, Thompson HW. Corneal thickness measurements with the Orbscan Topography System and ultrasonic pachymetry. J Cataract Refract Surg. 1997; 23:1345–1350.
22. Kang PS, Yang YS, Kim JD. Comparison of corneal thickness measurements with the orbscan and ultrasonic pachymetry. J Korean Ophthalmol Soc. 2000; 41:1697–1703.
23. Yeter V, Sönmez B, Beden U. Comparison of central corneal thickness measurements by Galilei Dual-Scheimpflug analyzer(R) and ultrasound pachymeter in myopic eyes. Ophthalmic Surg Lasers Imaging. 2012; 43:128–134.
24. Ladi JS, Shah NA. Comparison of central corneal thickness measurements with the Galilei dual Scheimpflug analyzer and ultrasound pachymetry. Indian J Ophthalmol. 2010; 58:385–388.
25. Lee MJ, Shin YU, Lim HW, et al. Central corneal thickness measured by noncontact specular microscopy, dual rotating scheimpflug camera and ultrasound pachymetry. J Korean Ophthalmol Soc. 2015; 56:1520–1526.
26. Maruyama Y, Mori K, Ikeda Y, et al. Effects of long-term topical prostaglandin therapy on central corneal thickness. J Ocul Pharmacol Ther. 2014; 30:440–444.
27. Sen E, Nalcacioglu P, Yazici A, et al. Comparison of the effects of latanoprost and bimatoprost on central corneal thickness. J Glaucoma. 2008; 17:398–402.
28. Grueb M, Mielke J, Rohrbach JM, Schlote T. Effect of brimonidine on corneal thickness. J Ocul Pharmacol Ther. 2011; 27:503–509.
29. Lee SY, Lee H, Bae HW, et al. Tear lipid layer thickness change and topical anti-glaucoma medication use. Optom Vis Sci. 2016; 93:1210–1217.
30. Herse P, Siu A. Short-term effects of proparacaine on human corneal thickness. Acta Ophthalmol (Copenh). 1992; 70:740–744.
31. Nam SM, Lee HK, Kim EK, Seo KY. Comparison of corneal thickness after the instillation of topical anesthetics: proparacaine versus oxybuprocaine. Cornea. 2006; 25:51–54.
32. Fukuda R, Usui T, Miyai T, et al. Corneal thickness and volume measurements by swept source anterior segment optical coherence tomography in normal subjects. Curr Eye Res. 2013; 38:531–536.
33. Kim IG, Lee CE, Lee JS, et al. Utility of the swept source optical coherence tomography for measurements of central corneal thickness. J Korean Ophthalmol Soc. 2016; 57:1542–1548.
Full Text Links
  • JKOS
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr