J Lab Med Qual Assur.  2011 Dec;33(2):89-93.

Interpretation of Weak-Positive Bands in a Multiplex PCR Using Seeplex RV12 ACE Detection Kit

Affiliations
  • 1Department of Laboratory Medicine, Hallym University College of Medicine, Chuncheon, Korea. kimhan@hallym.ac.kr

Abstract

BACKGROUND
Multiplex PCR assay is a sensitive tool for the detection of various respiratory viruses. Seeplex RV12 ACE Detection (Seegene, Korea) assay is a multiplex RT-PCR assay for detection of 12 respiratory viruses. We had observed several cases with faint bands in the test. Those results were investigated in this study.
METHODS
A total of 163 specimens were tested with Seeplex RV12 ACE Detection assay. The amplicons showing faint band in electrophoresis were reamplified and sequenced.
RESULTS
A total of 99 viruses were detected in 80 specimens (49.1%). Twenty-four amplicons showed faint band in eletrophoresis. All of influenza virus A, parainfluenza viruses (PIV), coronavirus OC43, human metapneumovirus (HMPV) and adenovirus amplicons were reamplified, but 4 of 12 human rhinovirus amplicons were not reamplified. Sequences of reamplified PCR products showed homology of 95-99% to those of corresponding viruses in the National Center for Biotechnology Information database.
CONCLUSIONS
Faint bands of influenza virus A, PIV-1, PIV-3, coronavirus OC43, HMPV and adenovirus in Seeplex RV12 ACE Detection assay are specific bands and seems to be weak positive results.

Keyword

Respiratory virus; Multiplex PCR; Weak positive

MeSH Terms

Adenoviridae
Biotechnology
Coronavirus
Coronavirus OC43, Human
Electrophoresis
Humans
Metapneumovirus
Multiplex Polymerase Chain Reaction
Orthomyxoviridae
Paramyxoviridae Infections
Polymerase Chain Reaction
Rhinovirus
Full Text Links
  • JLMQA
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr