Lab Med Online.  2015 Oct;5(4):188-195. 10.3343/lmo.2015.5.4.188.

Evaluation of the Automated Hematology Analyzer Sysmex XN-2000 and the Accuracy of Differential Leukocyte Counts Using the Low WBC Mode

Affiliations
  • 1Department of Laboratory Medicine, Busan Paik Hospital, Inje University College of Medicine, Busan, Korea.
  • 2Department of Laboratory Medicine, Haeundae Paik Hospital, Inje University College of Medicine, Busan, Korea. jeong418@medimail.co.kr

Abstract

BACKGROUND
The XN-series (Sysmex, Japan) is the new hematology analyzer from Sysmex, with new channels to improve the accuracy of differential leukocyte count and platelet count in the low cell count range. We evaluated the analytical performance and low white blood cell (WBC) mode of the XN-2000.
METHODS
Precision, linearity, and carryover were evaluated for the analyzer. We analyzed the accordance of complete blood count (CBC), reticulocyte count, and differential leukocyte count between the XN-2000 and XE-2100 (Sysmex), using 200 samples from normal controls and patients. For 80 samples with a WBC count <1.5x10(9) cells/L, the low WBC mode was evaluated by comparing the automated count with a manual differential count as the reference.
RESULTS
The coefficients of variation of precision were <5% for most CBC parameters and <10% for differential leukocyte count. All results obtained with the XN-2000 showed good correlation with those obtained with the XE-2100. The correlation coefficients (r) were >0.9800 for all CBC parameters except mean corpuscular hemoglobin concentration, mean platelet volume, and platelet distribution width, and >0.9900 for differential leukocyte count except monocytes and basophils. The low WBC mode provided accurate counts for neutrophils and lymphocytes, with r>0.9300 for samples with a WBC count of 0.1-1.5x10(9) cells/L.
CONCLUSIONS
The XN-2000 showed good analytical performance and correlation with the existing model, the XE-2100. The XN-2000 provided accurate results for differential leukocyte count in samples with a WBC count of 0.1-1.5x10(9) cells/L, and reduced manual slide reviews.

Keyword

Hematology analyzer; XN-2000; Differential leukocyte count; Low cell count

MeSH Terms

Basophils
Blood Cell Count
Blood Platelets
Cell Count
Erythrocyte Indices
Hematology*
Humans
Leukocyte Count*
Leukocytes
Lymphocytes
Mean Platelet Volume
Monocytes
Neutrophils
Platelet Count
Reticulocyte Count

Figure

  • Fig. 1. Linearities of white blood cell (WBC) count, red blood cell (RBC) count, hemoglobin (Hb), hematocrit (HCT), and platelet (PLT) count determined by the XN-2000 (A, XN-10 B3; B, XN-10 B4).

  • Fig. 2. Comparison of complete blood count parameters including mean platelet volume (MPV) and platelet distribution width (PDW) between the XN-2000 and XE-2100 using Bland-Altman plots (A, XN-10 B3; B, XN-10 B4).


Reference

References

1. Ward PC. The CBC at the turn of the millennium: an overview. Clin Chem. 2000; 46:1215–20.
Article
2. Buttarello M and Plebani M. Automated blood cell counts: state of the art. Am J Clin Pathol. 2008; 130:104–16.
3. Meintker L, Ringwald J, Rauh M, Krause SW. Comparison of automated differential blood cell counts from Abbott Sapphire, Siemens Advia 120, Beckman Coulter DxH 800, and Sysmex XE-2100 in normal and pathologic samples. Am J Clin Pathol. 2013; 139:641–50.
Article
4. Briggs C, Longair I, Kumar P, Singh D, Machin SJ. Performance evaluation of the Sysmex haematology XN modular system. J Clin Pathol. 2012; 65:1024–30.
Article
5. International Council for Standardization in Haematology. Guidelines for the evaluation of blood cell analysers including those used for differential leucocyte and reticulocyte counting and cell marker applications. International Council for Standardization in Haematology: prepared by the ICSH Expert Panel on Cytometry. Clin Lab Haematol. 1994; 16:157–74.
6. Clinical and Laboratory Standards Institute. Validation, verification, and quality assurance of automated hematology analyzers: approved standard. 2nd ed.CLSI document H26-A2. Wayne PA: Clinical and Laboratory Standard Institute;2010.
7. Clinical and Laboratory Standards Institute. Measurement procedure comparison and bias estimation using patient samples: approved guideline. 3rd ed.CLSI document EP9-A3. Wayne PA: Clinical and Laboratory Standards Institute;2013.
8. Clinical and Laboratory Standards Institute. Reference leukocyte (WBC) differential count(proportional) and evaluation of instrumental methods: approved standard. 2nd ed.CLSI document H20-A2.Wayne PA: Clinical and Laboratory Standard Institute;2007.
9. Seo JY, Lee ST, Kim SH. Performance evaluation of the new hematology analyzer Sysmex XN-series. Int J Lab Hematol. 2015; 37:155–64.
Article
10. Hove LV, Schisano T, Brace L. Anemia diagnosis, classification, and monitoring using Cell-Dyn technology reviewed for the new millennium. Lab Hematol. 2000; 6:93–108.
11. Leader A, Pereg D, Lishner M. Are platelet volume indices of clinical use? A multidisciplinary review. Ann Med. 2012; 44:805–16.
Article
12. Jo YA, Kim M, Kim HS, Kang HJ, Lee YK. Evaluation of the Mindray BC-6800 Complete Blood Counts Analyzer. Lab Med Online. 2013; 3:131–7.
Article
13. Kang SH, Shin E, Han CK, Kim HK, Cho HI. CELL-DYN Sapphire hematology analyzer performance evaluation on leukocyte differential count. J Lab Med Qual Assur. 2006; 28:219–24.
14. Park Y, Song J, Song S, Song KS, Ahn MS, Yang MS, et al. Evaluation of the Abbott Cell-Dyn Sapphire hematology analyzer. Korean J Lab Med. 2007; 27:162–8.
Article
15. Lesesve JF, Benbih M, Lecompte T. Accurate basophils counting: not an easy goal! Clin Lab Haematol. 2005; 27:143–4.
Full Text Links
  • LMO
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr