Korean J Obes.
2007 Jun;16(2):50-57.
Comparison of Bioelectrical Impedance Analysis with Dual Energy X-ray Absorptiometry in Obese Women
- Affiliations
-
- 1Department of Family Medicine, Seoul Paik Hospital, College of Medicine, Inje University, Korea. fmleader@nuri.net
- 2Obesity Research Institute of Inje University, Korea.
Abstract
-
BACKGROUND: Bioelectrical impedance analysis (BIA) is simple, inexpensive, portable, safe and relatively accurate means of assessing body composition. However, there were many reports in that the error of BIA was increased in extremely obese subjects. This study was conducted to assess the accuracy of BIA in obese women and to investigate factors that influence the error of BIA in obese women.
METHODS
173 healthy obese women (ages 21 to 49 years; BMI > or = 25 kg/m2) underwent BIA and dual energy X-ray absorptiometry (DEXA). The accuracy of BIA was assessed by Pearson correlation and Bland-Altman analysis. Factors that influence the error of BIA were investigated by Pearson correlation, one-way ANOVA, and multiple regression analysis.
RESULTS
In obese women, the correlations between estimates of BIA and DEXA were highly significant (r = 0.838~0.910). Compared with DEXA, BIA was highly accurate for fat mass [deltaFM (BIA-DEXA): 0.5 kg (95% confidence interval: -3.8~4.8)], but BIA overestimated fat free mass [deltaFM (BIA-DEXA): 2.8 kg (-1.9~7.5; 95% confidence interval)]. As adiposity was increasing, BIA overestimated fat free mass and underestimated fat mass. Increasing of waist circumference decreased the error of BIA occurred by increasing of adiposity.
CONCLUSION
In obese women, the correlations and agreements between estimates of BIA and DEXA were highly significant. However, the amount and distribution of body fat had influence on the error of BIA.