Prog Med Phys.  2022 Dec;33(4):63-71. 10.14316/pmp.2022.33.4.63.

Effect of Bead Device Diameter on Z-Resolution Measurement in Tomosynthesis Images: A Simulation Study

Affiliations
  • 1Department of Radiological Technology, Faculty of Health Sciences, Okayama University, Okayama, Japan
  • 2Department of Radiological Technology, Okayama University Hospital, Okayama, Japan

Abstract

Purpose
To clarify the relationship between the diameter of the simulated bead and the Z-resolution of the tomosynthesis image.
Methods
A simulated bead was placed on a 1,024×1,024×1,024-pixel base image. The diameters were set to 0.025, 0.05, 0.1, 0.2, 0.3, 0.7, 1.0, and 1.3 mm. A bead was placed at the center of the base image and projected at a simulated X-ray angle range of ±45° to obtain a projected image. A region of interest was placed at the center of the bead image and the slice sensitivity profile (SSP) was obtained by acquiring pixel values in the z-direction. The full width at half maximum of the SSP was defined as the Z-resolution and the frequency response was obtained by the 1-D Fourier transform of the SSP.
Results
Z-resolution increased with increasing bead diameter. However, there was no change in Z-resolution between 0.025 and 0.1 mm. The frequency response was similar to that of the Z-resolution, with a significant difference between 0.1 and 0.2 mm diameter.
Conclusions
Z-resolution is dependent on the diameter of the bead, which should be selected considering the pixel size of the tomosynthesis image.

Keyword

Tomosynthesis; Z-resolution; Frequency response; Bead method

Figure

  • Fig. 1 Simulated bead images. The pixel values of the base image and supporter were set to 0 and 60, respectively.

  • Fig. 2 Each simulated bead image. The pixel value of the bead image was 200; however, for a diameter of 0.025 mm, it was half (100) than that of the other diameters.

  • Fig. 3 Schematic of the virtual projection for the projected image. The number of pixels of the focus-shift value and the minimum D value employed are 1,000 and 500 pixels, respectively. Therefore, the X-ray tube angular range is ±45°.

  • Fig. 4 Simulated projected images. The number of projected images was 74.

  • Fig. 5 Low-pass filter shapes applied to the filtered back-projection.

  • Fig. 6 Simulated (a) and real (b) tomosynthesis images of the bead. The numbers below the figures correspond to the simulated diameter.

  • Fig. 7 Schematic of the acquisition for the real projected image using general X-ray equipment. The angular range of the real bead image (±30°) differed from the simulated bead image (±45°). However, the apparent angular range to the bead increases as the X-ray tube moves around the table surface.

  • Fig. 8 Schematic of the Z-resolution measurement method. A region of interest was placed at the center of the bead image, and slice sensitivity profile was acquired by plotting the maximum pixel values of the bead images. FWHM, full width at half maximum.

  • Fig. 9 Acquired slice sensitivity profile of the simulated and real bead image. The Z-resolutions of the simulated and real bead image were 2.33 and 2.68 mm, respectively.

  • Fig. 10 Acquired slice sensitivity profile (SSP) of each simulated bead image. The shape of SSP of 0.025 mm was equivalent to these of the 0.05 and 0.1 mm.

  • Fig. 11 Results of the Z-resolution of each simulated bead diameter. There were no change of the Z-resolution value of the diameter of 0.025, 0.05, and 0.1 mm.

  • Fig. 12 Results of the frequency response of each simulated bead diameter. The frequency responses of 0.2 and 0.3 mm were diverged comparison with the results of the Z-resolution.

  • Fig. 13 The x-z plane view of simulated bead of (a) 0.025 mm, (b) 0.05 mm, and (c) 0.1 mm.


Reference

References

1. Yaffe MJ. 2014; Research in digital mammography and tomosynthesis at the University of Toronto. Radiol Phys Technol. 7:191–202. DOI: 10.1007/s12194-014-0277-y. PMID: 24961727.
Article
2. Johnsson AA, Vikgren J, Bath M. 2014; Chest tomosynthesis: technical and clinical perspectives. Semin Respir Crit Care Med. 35:17–26. DOI: 10.1055/s-0033-1363448. PMID: 24481756.
Article
3. Livingston KS, Edwards EA, Griffin M, MacKenzie JD, Zapala MA. 2021; Use of digital tomosynthesis in assessing accurate medial epicondyle fracture displacement as compared with conventional radiography and computed tomography. J Pediatr Orthop. 41:e877–e883. DOI: 10.1097/BPO.0000000000001917. PMID: 34419980. PMCID: PMC8508716.
Article
4. Hu YH, Zhao B, Zhao W. 2008; Image artifacts in digital breast tomosynthesis: investigation of the effects of system geometry and reconstruction parameters using a linear system approach. Med Phys. 35:5242–5252. DOI: 10.1118/1.2996110. PMID: 19175083. PMCID: PMC2673612.
Article
5. Zheng J, Fessler JA, Chan HP. 2019; Effect of source blur on digital breast tomosynthesis reconstruction. Med Phys. 46:5572–5592. DOI: 10.1002/mp.13801. PMID: 31494953. PMCID: PMC6899200.
Article
6. van Engen RE, Bosmans H, Bouwman RW, Dance DR, Heid P, Lazzari B, et al. 2018. Protocol for the quality control of the physical and technical aspects of digital breast tomosynthesis systems (version 1. EUREF;Berlin: p. 30–44. DOI: 10.1055/s-0033-1363448.
7. Fukui R, Ishii R, Kishimoto J, Yamato S, Takahata A, Kohama C. 2014; Evaluation of the effect of geometry for measuring section thickness in tomosynthesis. Radiol Phys Technol. 7:141–147. DOI: 10.1007/s12194-013-0243-0. PMID: 24254729.
Article
8. Maldera A, De Marco P, Colombo PE, Origgi D, Torresin A. 2017; Digital breast tomosynthesis: dose and image quality assessment. Phys Med. 33:56–67. DOI: 10.1016/j.ejmp.2016.12.004. PMID: 28010921.
Article
9. Sage J, Fezzani KL, Fitton I, Hadid L, Moussier A, Pierrat N, et al. 2019; Experimental evaluation of seven quality control phantoms for digital breast tomosynthesis. Phys Med. 57:137–144. DOI: 10.1016/j.ejmp.2018.12.031. PMID: 30738517.
Article
10. Mertelmeier T, Orman J, Haerer W, Dudam MK. 2006. Optimizing filtered backprojection reconstruction for a breast tomosynthesis prototype device. Paper presented at: Medical Imaging 2006: Physics of Medical Imaging; 2006 Feb 11-16; San Diego, USA. DOI: 10.1117/12.651380.
Article
Full Text Links
  • PMP
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr