J Korean Med Sci.  2019 Dec;34(49):e314. 10.3346/jkms.2019.34.e314.

Comparison of Operator Workloads Associated with the Single-unit Anyfusion® Pump and the Changeover from a Syringe Pump to an Infusion Pump

Affiliations
  • 1Department of Anesthesiology and Pain Medicine, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Korea. dikei@hanmail.net

Abstract

BACKGROUND
A changeover from a syringe pump to an infusion pump may transiently interrupt drug administration and result in medication errors. A newly developed infusion device (Anyfusion®) that combines the functions of an infusion pump and a syringe pump in a single unit was recently commercialized. This study compared the operator workloads associated with the Anyfusion® pump and with the conventional method using two infusion devices (a syringe pump and an infusion pump).
METHODS
This was a prospective, randomized, crossover-designed simulation study using an imitation arm with an intravascular line. We compared the two methods in terms of total execution time, total number of button clicks, and subjective difficulty using a numerical rating scale with a score from 0 (extremely easy) to 10 (extremely difficult).
RESULTS
Twenty-two nurses successfully performed both interventions according to the allocated sequence. Total execution times did not differ between the two methods (129.5 ± 23.2 seconds for the conventional method vs. 121.2 ± 24.3 seconds for the Anyfusion® method; P = 0.244), although the total number of clicks was significantly fewer using the Anyfusion® than the conventional method (median [interquartile range]: 10.0 [9.0-12.0] vs. 21.0 [20.0-25.0], respectively; P < 0.001). Participants rated the Anyfusion® method as easier than the conventional method (1.7 ± 1.2 vs. 3.6 ± 1.6; P < 0.001, respectively).
CONCLUSION
The introduction of Anyfusion® lessened the workload of practitioners required by the changeover process, which might reduce the risk of medication errors and subsequent patient harm. TRIAL REGISTRATION: Clinical Research Information Service Identifier: KCT0004172

Keyword

Infusion Pumps; Medication Errors; Workload

MeSH Terms

Arm
Information Services
Infusion Pumps*
Medication Errors
Methods
Patient Harm
Prospective Studies
Syringes*
Full Text Links
  • JKMS
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr