Lab Med Online.  2019 Jul;9(3):133-145. 10.3343/lmo.2019.9.3.133.

Evaluation of the Analytical Performance of Atellica CH 930 Automated Chemistry Analyzer

Affiliations
  • 1Department of Laboratory Medicine, Yonsei University College of Medicine, Seoul, Korea. comforter6@yuhs.ac
  • 2Department of Laboratory Medicine, Wonju Severance Christian Hospital, Yonsei University Wonju College of Medicine, Wonju, Korea.

Abstract

BACKGROUND
Recently, a new automated chemistry analyzer, Atellica CH930 (Siemens, Germany), was introduced. It automatically measures internal quality control (QC) materials according to a pre-determined schedule. For this purpose, the instrument has space for storage of QC materials. We evaluated the analytical performance of chemistry items by using the Atellica system.
METHODS
The precision of 29 items was evaluated with three levels of QC materials with two storage methods. We stored the QC materials in the dedicated storage space in the instrument during the precision evaluation period. In addition, we aliquoted and stored the materials in the refrigerator, and then loaded the material in a timely manner. Linearity, carry-over, and agreement with current methods were also evaluated.
RESULTS
The within-laboratory coefficient of variation (CV) of most items, except for total CO2 (tCO2), was within 5.0% in both QC storage methods without significant differences in CV between storage methods. The CV of tCO2 was 5.2%, 5.8%, and 5.1% at three different levels when the QC materials were stored in a dedicated space in the instrument. The linearity was acceptable, showing <5% nonlinearity. Although good agreement was observed for most items, some items, such as calcium, total bilirubin, aspartate transaminase, and chloride, showed unequivalent results.
CONCLUSIONS
Atellica CH930 showed acceptable precision, linearity, and agreement in routine chemistry items. The automatic QC function using the storage device has no problem with stability or precision. It can reduce the manual process, allowing technicians to focus on reviewing the QC results and reporting reliable results.

Keyword

Atellica CH 930 Analyzer; Performance evaluation; Automated QC function

MeSH Terms

Appointments and Schedules
Aspartate Aminotransferases
Bilirubin
Calcium
Chemistry*
Quality Control
Aspartate Aminotransferases
Bilirubin
Calcium

Figure

  • Fig. 1. Precision of the Siemens Atellica CH 930 Analyzer. Within-laboratory precision (A) and repeatability (B) of 29 analytes were calculated according to the CLSI guideline EP05-A3. QC materials used were supplied either by conventional manual loading after warming up to room temperature from refrigerator storage (black) or by automatic loading from the inner storage/supply system of the instrument (red). Mean CV value and the 95% confidence interval of each analyte are denoted as rectangles and flanking lines, respectively.

  • Fig. 2. Agreement and correlation of 28 chemical analytes between the Atellica CH 930 analyzer (Y axis) and Hitachi 7600 analyzer (X axis).

  • Fig. 3. Agreement and correlation of 25 chemical analytes between the Atellica CH 930 analyzer (Y axis) and Cobas c702 (X axis).


Reference

References

1. Brombacher PJ, Marell GJ, Westerhuis LW. Laboratory work fow analysis and introduction of a multifunctional analyser. Eur J Clin Chem Clin Biochem. 1996; 34:287–92.
2. Armbruster DA, Overcash DR, Reyes J. Clinical chemistry laboratory automation in the 21st century – Amat Victoria curam (Victory loves careful preparation). Clin Biochem Rev. 2014; 35:143–53.
3. Dolci A, Giavarina D, Pasqualetti S, Szőke D, Panteghini M. Total laboratory automation: Do stat tests still matter? Clin Biochem. 2017; 50:605–11.
Article
4. Lawson NS, Haven GT, Williams GW. Analyte stability in clinical chemistry quality control materials. Crit Rev Clin Lab Sci. 1982; 17:1–50.
Article
5. Clinical and Laboratory Standards Institute. Evaluation of precision of quantitative measurement procedures; approved guideline—third edition. CLSI document EP05-A3. Wayne, PA: Clinical and Laboratory Standards Institute;2014.
6. Clinical and Laboratory Standards Institute. Evaluation of the of the linearity of quantitative measurement procedures: a statistical approach; approved guideline. CLSI document EP06-A. Wayne, PA: Clinical and Laboratory Standards Institute;2003.
7. Clinical and Laboratory Standards Institute. Method comparison and bias estimation using patient samples; approved guideline—second edition (interim revision). CLSI document EP09-A2-IR. Wayne, PA: Clinical and Laboratory Standards Institute;2010.
8. Statland BE, ed. Clinical decision levels for laboratory tests. 2nd ed.Oradell, NJ: Medical Economics Books;1987.
9. Burtis CA, Ashwood ER, et al. eds. Tietz textbook of clinical chemistry and molecular diagnostics. 4th ed.St. Louis: Elsevier-Saunders;2006. p. 363–4.
10. Grundy SM, Becker D, Clark LT, Cooper RS, Denke MA, Howard J, et al. Detection, evaluation, and treatment of high blood cholesterol in adults (Adult Treatment Panel III). Circulation. 2002; 106:3143–421.
11. Ricos C, Alvarez V, Cava F, Garcia-Lario JV, Hernandez A, Jimenez CV, et al. Desirable specifcations for total error, imprecision, and bias, derived from intra- and interindividual biologic variation. https://www.westgard.com/biodatabase1.htm. (Updated in 2014).
12. Westgard QC. CLIA requirements for analytical quality. https://www.westgard.com/clia.htm. (Updated on Feb, 1992).
13. Burtis CA, Ashwood ER, et al. eds. Tietz textbook of clinical chemistry and molecular diagnostics. 5th ed.Philadelphia: Elsevier-Saunders;2014.
14. Westgard QC. Royal College of Pathologists of Australasia analytical quality requirements. https://www.westgard.com/rcpa-biochemistry. htm#serumchemistry (Updated in 2013).
15. Greg Miller W, Myers GL, Lou Gantzer M, Kahn SE, Schönbrunner ER, Thienpont LM, et al. Roadmap for harmonization of clinical laboratory measurement procedures. Clin Chem. 2011; 57:1108–17.
Article
16. Klauke R, Kytzia HJ, Weber F, Grote-Koska D, Brand K, Schumann G. Reference measurement procedure for total bilirubin in serum re-evaluated and measurement uncertainty determined. Clin Chim Acta. 2018; 481:115–20.
Article
17. Lo SF, Kytzia HJ, Schumann G, Swartzentruber M, Vader HL, Weber F, et al. Interlaboratory comparison of the Doumas bilirubin reference method. Clin Biochem. 2009; 42:1328–30.
Article
18. Lo S, Jendrzejczak B, Doumas BT. Bovine serum–based bilirubin calibrators are inappropriate for some diazo methods. Clin Chem. 2010; 56:869–72.
Article
19. Infusino I, Frusciante E, Braga F, Panteghini M. Progress and impact of enzyme measurement standardization. Clin Chem Lab Med. 2017; 55:334–40.
Article
Full Text Links
  • LMO
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr