Korean J Thorac Cardiovasc Surg.
2000 Feb;33(2):146-150.
transprosthetic Pressure Gradient after aortic Valve Replacement with Small
Sized Prostheses
- Affiliations
-
- 1Department of Thoracic and Cardiovascular Surgery, Dong Kwang Ju Hospital.
- 2Department of Thoracic and Cardiovascular Surgery, Samsung Medical Center, Sungkyunkwan University School of Medicine.
Abstract
-
BACKGROUND: The prognosis after an aortic valve replacment can be affected significantly by
the transprosthetic pressure gradient which is determined mainly by the size of the patients
body and the prosthesis used. We analyzed the hemodynamic feature of two relatively new
prosthese the ATS and the evensized Medtronic-Hall(M-H) valves by measuring the
transprosthetic pressure gradient in the cases where small sizes (23mm or smaller) were used.
MATERIAL AND METHOD: There were 94 patients who received whom aortic valve replacement with
prosthesis smaller than 23 mm from October 1994 to June 1998. In these patients the
transprosthetic pressure gradient clalculated from the pressure half time during postoperative
Dopper echocardiographic examination was compared between the prostheses of different sizes.
The body surface area of each patient was also taken into consideration.
RESULT: The mean pressure gradient and body surface area in each group were 21.7+/-10.2 mmHg
and 1.52+/-0.14m2 in ATS 19mm 11.4+/-6.5 mmHg and 1,57+/-0.20m2 in M-H 20mm 15.2+/-6.3 mmHg and 1.54+/-0.13m2 in ATS 21mm 9.3+/-2.5 mmHg and 1.63 +/-0.14m2 in M-H 22 mm and 12.9+/-5.3 mmHg and 1.69+/-0.13m2 in ATS 23mm.
CONCLUSIONS
The 19mm ATS prosthesis showed significant trasprosthetic pressure gradient which
is similar to the values previously reported with other bileaflet prosthesesm Close follow-up
was needed in terms of exercise capacity and change in left ventiricular geometry. In patients
with small aortic valve annulus the 20mm M-H valve is recomendable as an alternative to 19mm
bileaflet valves because it has less pressure gradient with similar outer diameter.