Korean J Med Phys.
2012 Jun;23(2):114-122.
Practical Output Dosimetry with Undefined NdwCo-60 of Cylindrical Ionization Chamber for High Energy Photon Beams of Linear Accelerator
- Affiliations
-
- 1Department of Radiation Oncology, School of Medicine, Keimyung University, Daegu, Korea.
- 2Department of Medical Engineering and Physics in Medicine, School of Medicine, Keimyung University, Daegu, Korea. tjchoi@dsmc.or.kr
- 3Department of Radiation Oncology, Chonnam National University Hospital, Gwangju, Korea.
Abstract
- For the determination of absorbed dose to water from a linear accelerator photon beams, it needs a exposure calibration factor Nx or air kerma calibration factor Nk of air ionization chamber. We used the exposure calibration factor Nx to find the absorbed dose calibration factors of water in a reference source through the TG-21 and TRS-277 protocol. TG-21 used for determine the absorbed dose in accuracy, but it required complex calculations including the chamber dependent factors. The authors obtained the absorbed dose calibration factor NdwCo-60 for reduce the complex calculations with unknown Ndw only with Nx or Nk calibration factor in a TM31010 (S/N 1055, 1057) ionization chambers. The results showed the uncertainty of calculated Ndw of IC-15 which was known the Nx and Ndw is within -0.6% in TG-21, but 1.0% in TRS-277. and TM31010 was compared the Ndw of SSDL to that of PSDL as shown the 0.4%, -2.8% uncertainty, respectively. The authors experimented with good agreement the calculated Ndw is reliable for cross check the discrepancy of the calibration factor with unknown that of TM31010 and IC-15 chamber.