Yeungnam Univ J Med.  1988 Dec;5(2):31-36. 10.12701/yujm.1988.5.2.31.

A Study on ARC Therapy of 18MeV Linear Accelerator

Abstract

In recent years there has been a growing interest in all forms of rotational therapy, and many different types of therapy machines designed for this kind of treatment have become available. To the medical radiation physicist, the dosimetry of rotation therapy has presented a number of interesting problems, and much useful work has been published on the basic date of dose distribution and dosage calculation. The setting dose for ARC therapy were obtained by computer calculation and measurement with cylindrical phantom. Authors compared computer calculation with measured value. And in ARC therapy, the region of maximum dose in shifted from the tumor center. The extent of shift was analyzed by isodose distribution for ARC therapy techniques.