J Korean Ophthalmol Soc.  2016 Aug;57(8):1228-1232. 10.3341/jkos.2016.57.8.1228.

Comparison of Partial Interferometry and Ultrasound A-scan for Axial Length Measurement in Retinal Vein Occlusions

Affiliations
  • 1Department of Ophthalmology, Chungnam National University School of Medicine, Daejeon, Korea. kimjy@cnu.ac.kr
  • 2Research Institute for Medical Sciences, Chungnam National University, Daejeon, Korea.

Abstract

PURPOSE
To evaluate the significance of axial length, which is a known risk factor of retinal vein occlusion, we measured the axial lengthby using contact A-scan sonography and partial interferometry and compared the two values.
METHODS
This study included 19 patients complaining of visual symptoms and who were diagnosed with unilateral retinal vein occlusion (RVO) with macular edema (ME). Affected eyes were classified as the study group, and healthy fellow eyes were classified as the control group. We measured the central macular thickness (CMT) and axial length (AL) of the affected and fellow eyes and compared them. CMT was measured by optical coherence tomography (Stratus OCT, Carl Zeiss, Jena, Germany), and AL was measured by interferometry (IOL Master®, Carl Zeiss, Jena, Germany).
RESULTS
In RVO patients, CMT was significantly different between affected eyes (485.7 ± 111.3 µm) and fellow eyes (197.8 ± 29.7 µm; p < 0.001). Axial length measured by A-scan sonography was 23.06 ± 0.86 mm in the affected eyes and 23.28 ± 0.93 mm in the healthy eyes, which was statistically different (p < 0.001). However, using partial interferometry, the AL was 23.35 ± 0.87 mm in the affected eyes and 23.38 ± 0.95 mm in the healthy eyes. No significant difference was found.
CONCLUSIONS
We confirmed that short AL, which was once thought to be a risk factor of RVO, results from the properties of the instruments used for measurement. Moreover, we verified that partial interferometry is more accurate for measurement of AL than A-scan sonography when retinal vein occlusion is associated with ME.

Keyword

Axial length; Central macular thickness; Macular edema; Retinal vein occlusion

MeSH Terms

Humans
Interferometry*
Macular Edema
Retinal Vein Occlusion*
Retinal Vein*
Retinaldehyde*
Risk Factors
Tomography, Optical Coherence
Ultrasonography*
Retinaldehyde

Reference

References

1. Hayreh SS, Zimmerman MB, Podhajsky P. Incidence of various types of retinal vein occlusion and their recurrence and abdominal characteristics. Am J Ophthalmol. 1994; 117:429–41.
2. Risk factors for branch retinal vein occlusion. The Eye Disease Case-control Study Group. Am J Ophthalmol. 1993; 116:286–96.
3. Lahey JM, Tunç M, Kearney J, et al. Laboratory evaluation of abdominal states in patients with central retinal vein occlusion who are less than 56 years of age. Ophthalmology. 2002; 109:126–31.
4. Lahey JM, Kearney JJ, Tunc M. Hypercoagulable states and abdominal retinal vein occlusion. Curr Opin Pulm Med. 2003; 9:385–92.
5. Cahill MT, Stinnett SS, Fekrat S. Meta-analysis of plasma abdominal, serum folate, serum vitamin B(12), and thermolabile MTHFR genotype as risk factors for retinal vascular occlusive disease. Am J Ophthalmol. 2003; 136:1136–50.
6. Aritürk N, Oge Y, Erkan D, et al. Relation between retinal vein abdominals and axial length. Br J Ophthalmol. 1996; 80:633–6.
7. Timmerman EA, de Lavalette VW, van den Brom HJ. Axial length as a risk factor to branch retinal vein occlusion. Retina. 1997; 17:196–9.
Article
8. Simons BD, Brucker AJ. Branch retinal vein occlusion. Axial length and other risk factors. Retina. 1997; 17:191–5.
9. Kumar A, Ahuja S, Tewari HK, et al. Short axial length: risk abdominals for central retinal vein occlusion. Ann Ophthalmol. 1994; 26:145–7.
10. Olsen T. The accuracy of ultrasonic determination of axial length in pseudophakic eyes. Acta Ophthalmol (Copenh). 1989; 67:141–4.
Article
11. Binkhorst RD. The accuracy of ultrasonic measurement of the abdominal length of the eye. Ophthalmic Surg. 1981; 12:363–5.
12. Giers U, Epple C. Comparison of A-scan device accuracy. J Cataract Refract Surg. 1990; 16:235–42.
Article
13. Vogel A, Dick HB, Krummenauer F. Reproducibility of optical biometry using partial coherence interferometry: intraobserver and interobserver reliability. J Cataract Refract Surg. 2001; 27:1961–8.
Article
14. Tehrani M, Krummenauer F, Kumar R, Dick HB. Comparison of biometric measurements using partial coherence interferometry and applanation ultrasound. J Cataract Refract Surg. 2003; 29:747–52.
Article
15. Brown MM, Brown GC, Menduke H. Central retinal vein abdominal and axial length. Ophthalmic Surg. 1990; 21:623–4.
16. Holzer MP, Mamusa M, Auffarth GU. Accuracy of a new partial coherence interferometry analyser for biometric measurements. Br J Ophthalmol. 2009; 93:807–10.
Article
17. Rose LT, Moshegov CN. Comparison of the Zeiss IOLMaster and applanation A-scan ultrasound: biometry for intraocular lens calculation. Clin Experiment Ophthalmol. 2003; 31:121–4.
Article
18. Connors R 3rd, Boseman P 3rd, Olson RJ. Accuracy and abdominal of biometry using partial coherence interferometry. J Cataract Refract Surg. 2002; 28:235–8.
19. Haigis W, Lege B, Miller N, Schneider B. Comparison of abdominal ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to Haigis. Graefes Arch Clin Exp Ophthalmol. 2000; 238:765–73.
20. Kovács I, Ferencz M, Nemes J, et al. Intraocular lens power abdominal for combined cataract surgery, vitrectomy and peeling of abdominal membranes for macular oedema. Acta Ophthalmol Scand. 2007; 85:88–91.
21. Ueda T, Nawa Y, Hara Y. Relationship between the retinal abdominal of the macula and the difference in axial length. Graefes Arch Clin Exp Ophthalmol. 2006; 244:498–501.
Full Text Links
  • JKOS
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr