Lab Med Online.  2014 Oct;4(4):187-190. 10.3343/lmo.2014.4.4.187.

Performance Evaluation of the Syva EMIT Methotrexate Assay on the Toshiba 200FR NEO

Affiliations
  • 1Department of Laboratory Medicine, College of Medicine, Chungnam National University, Daejeon, Korea. kckwon@cnu.ac.kr
  • 2Department of Laboratory Medicine, Eulji University Hospital, Daejeon, Korea.

Abstract

BACKGROUND
Methotrexate (MTX) is an antifolate antagonist that is widely used for treating various malignancies and non-malignant diseases. MTX levels should be monitored when used in high concentration to determine when to start leucovorin rescue. In this study, we evaluated the analytical performance of the EMIT Methotrexate Assay on a 200FR NEO Chemistry Analyzer (Toshiba Medical System Co., Japan) and compared it with Viva-E Drug Testing System (Siemens Healthcare, Germany).
METHODS
According to the Clinical Laboratory and Standards Institute (CLSI) Evaluation Protocol (EP) 5-A2, three concentrations of the Liquichek Therapeutic Drug Monitoring Control (Bio-Rad Laboratories, USA) were analyzed twice a day for 20 days to monitor assay precision. The 200FR NEO and Viva-E instruments were compared using 40 patients' sera, according to CLSI EP9-A2. The linearity and carry-over rate were also evaluated.
RESULTS
Between-run CVs for low-, medium-, and high-level controls were 4.9%, 0.9%, and 2.0%, respectively, whereas between-day CVs for low-, medium-, and high-level controls were 8.1%, 1.3%, and 3.5%, respectively. In the linearity test, the coefficient of determination (R2) was 0.98 (0.06-1.92 micromol/L). In the comparison study, R2 was 0.955, showing good correlation between the 200FR NEO and Viva-E instruments. The carry-over rate was 0.9%.
CONCLUSIONS
The EMIT assay showed good precision, linearity, and carry-over rate on the Toshiba 200FR. An excellent correlation was observed when comparing results obtained using the Toshiba and Viva-E instruments. In conclusion, the Syva EMIT MTX assay can be readily used for MTX monitoring on the Toshiba 200FR NEO.

Keyword

Methotrexate; EMIT; Therapeutic drug monitoring

MeSH Terms

Chemistry
Delivery of Health Care
Drug Monitoring
Leucovorin
Methotrexate*
Leucovorin
Methotrexate

Figure

  • Fig. 1 Comparison of MTX level determinations measured by the Viva-E and the Toshiba 200FR NEO. Differences in results from either instrument are shown in (A) a scatter plot with an ordinary linear fit (blue line), (B) a Bland-Altman difference plot, or (C) a Bland-Altman percent difference plot.

  • Fig. 2 Linearity of MTX values using the Syva EMIT Methotrexate Assay on the Toshiba 200FR NEO.


Reference

1. Chung HS, Lee ST, Lee SY. Evaluation of Viva-E drug testing system. Korean J Lab Med. 2007; 27:330–337.
Article
2. Klee GG, Westgard JO. Quality management. In : Burtis CA, Ashwood ER, Bruns DE, editors. Tietz textbook of clinical chemistry and molecular diagnostics. 5th Ed. St. Louis: Elsevier Saunders;2012. p. 163–203.
3. Borgman MP, Hiemer MF, Molinelli AR, Ritchie JC, Jortani SA. Improved sensitivity for methotrexate analysis using enzyme multiplied immunoassay technique on the Siemens Viva-E instrument. Ther Drug Monit. 2012; 34:193–197.
Article
4. Clinical and Laboratory Standards Institute. CLSI document EP5-A2. Evaluation of precision performance of quantitative measurement methods: approved guideline. 2nd ed. Wayne: Clinical and Laboratory Standards Institute;2004.
5. Clinical and Laboratory Standards Institute. CLSI document EP9-A2. Method comparison and bias estimation using patient samples: approved guideline. 2nd ed. Wayne: Clinical and Laboratory Standards Institute;2002.
6. Clinical and Laboratory Standards Institute. CLSI document EP6-A2. Evaluation of the linearity of quantitative measurement procedures: a statistical approach. Approved guideline. Wayne: National Committee for Clinical Laboratory Standards;2003.
7. Glazko AJ. Phenytoin, chemistry and methods of determination. In : Levy RH, editor. Antiepileptic drugs. 3rd ed. New York: Raven press;1989. p. 159–176.
8. Steijns LS, Bouw J, van der Weide J. Evaluation of fluorescence polarization assays for measuring valproic acid, phenytoin, carbamazepine and phenobarbital in serum. Ther Drug Monit. 2002; 24:432–435.
Article
9. Armstrong VW, Oellerich M. Critical evaluation of methods for therapeutic drug monitoring. In : Evans WE, editor. Applied pharmacokinetics and pharmacodynamics. 4th ed. Baltimore: Lippincott Williams & Wilkins;2006. p. 30–39.
10. Mendu DR, Chou PP, Soldin SJ. An improved application for the enzyme multiplied immunoassay technique for caffeine, amikacin, and methotrexate assays on the Dade-Behring Dimension RxL Max clinical chemistry system. Ther Drug Monit. 2007; 29:632–637.
Article
11. Participant summary, external quality assessment in therapeutic drug monitoring and drug of abuse. Accessed October 2013. http://www.lab-qa.org.
Full Text Links
  • LMO
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr