Korean J Anesthesiol.  2012 Mar;62(3):234-239. 10.4097/kjae.2012.62.3.234.

Spectral entropy for assessing the depth of propofol sedation

Affiliations
  • 1Department of Anesthesiology, National Medical Center, Seoul, Korea.
  • 2Department of Anesthesiology, School of Medicine, Jeju National University, Jeju, Korea.
  • 3Department of Anesthesiology, Konkuk University Medical Center, Research Institute of Biomedical Science, Konkuk University School of Medicine, Seoul, Korea. taeyop@kuh.ac.kr

Abstract

BACKGROUND
For patients in the intensive care unit (ICU) or under monitored anesthetic care (MAC), the precise monitoring of sedation depth facilitates the optimization of dosage and prevents adverse complications from underor over-sedation. For this purpose, conventional subjective sedation scales, such as the Observer's Assessment of Alertness/Sedation (OAA/S) or the Ramsay scale, have been widely utilized. Current procedures frequently disturb the patient's comfort and compromise the already well-established sedation. Therefore, reliable objective sedation scales that do not cause disturbances would be beneficial. We aimed to determine whether spectral entropy can be used as a sedation monitor as well as determine its ability to discriminate all levels of propofol-induced sedation during gradual increments of propofol dosage.
METHODS
In 25 healthy volunteers undergoing general anesthesia, the values of response entropy (RE) and state entropy (SE) corresponding to each OAA/S (5 to 1) were determined. The scores were then analyzed during each 0.5 mcg/ml- incremental increase of a propofol dose.
RESULTS
We observed a reduction of both RE and SE values that correlated with the OAA/S (correlation coefficient of 0.819 in RE-OAA/S and 0.753 in SE-OAA/S). The RE and SE values corresponding to awake (OAA/S score 5), light sedation (OAA/S 3-4) and deep sedation (OAA/S 1-2) displayed differences (P < 0.05).
CONCLUSIONS
The results indicate that spectral entropy can be utilized as a reliable objective monitor to determine the depth of propofol-induced sedation.

Keyword

Entropy; Propofol; Sedation

MeSH Terms

Anesthesia, General
Deep Sedation
Entropy
Humans
Intensive Care Units
Light
Organothiophosphorus Compounds
Propofol
Weights and Measures
Organothiophosphorus Compounds
Propofol

Cited by  2 articles

Bispectral index monitoring correlates with the level of consciousness in brain injured patients
Jin Yong Jung, Cheol Beom Cho, Bo Mi Min
Korean J Anesthesiol. 2013;64(3):246-250.    doi: 10.4097/kjae.2013.64.3.246.

An unexpected increase of entropy in a sleepwalking disorder patient during propofol and remifentanil anesthesia: a case report
Yoon Ji Choi, Koo Kwon, Go Eun Bae, Seung Zhoo Yoon, Hye Won Lee, Hae Ja Lim
Korean J Anesthesiol. 2014;67(4):270-274.    doi: 10.4097/kjae.2014.67.4.270.

Full Text Links
  • KJAE
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr