Korean J Anesthesiol.  2007 May;52(5):578-581. 10.4097/kjae.2007.52.5.578.

Awareness during General Anesthesia due to Vaporizer Malfunction: A case report

Affiliations
  • 1Department of Anesthesiology and Pain Medicine, Seoul National University Hospital, Seoul National University College of Medicine, Korea. aws@snu.ac.kr
  • 2Department of Anesthesiology and Pain Medicine, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea.

Abstract

We report a case of awakening during general anesthesia due to a vaporizer malfunction. The sevoflurane vaporizer had a hole through which approximately 20% of fresh gas escaped. The gas in the common gas outlet contained only 60% of the sevoflurane stated on the vaporizer setting. Moreover, the gas monitor module was out of order, and the heart rate and blood pressure were stable. As a result, we were unaware of the low sevoflurane concentration. The leakage through the hole could not be detected with the commonly used low-pressure system leak checking method. The implication of this case is that unexpected awakening can occur in patients with stable vital signs with an inhalation anesthesia. Therefore, more attention is needed to detect the level of patient awareness.

Keyword

awareness; preanesthesia machine check; sevoflurane; vaporizer malfunction

MeSH Terms

Anesthesia, General*
Anesthesia, Inhalation
Blood Pressure
Heart Rate
Humans
Nebulizers and Vaporizers*
United Nations
Vital Signs
Full Text Links
  • KJAE
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr