Anesth Pain Med.  2013 Oct;8(4):237-239.

Interruption of fresh gas supply without alarm signs by the partial disconnection of fresh gas outlet during general anesthesia: A case report

Affiliations
  • 1Department of Anesthesiology and Pain Medicine, Korea University Medical Center, Anam Hospital, Seoul, Korea. drchois@korea.ac.kr

Abstract

We report our experience of unexpected low expiratory concentration of sevoflurane 0.8 vol%, low inspired fraction of oxygen 31%, and high value of bispctral index (BIS) 62 with sevoflurlane vaporizer setting 2.0 vol% and the inspired fraction of oxygen 50% during general anesthesia of a 56-year-old man undergoing elective laparoscopic hemicolectomy. The anesthesia machine (Cato edition, Drager, Lubeck, Germany) with a piston-based, electronically controlled ventilator was used for anesthesia. We checked thoroughly all the possible causes of the discrepancy of anesthetic gas concentration, and then detected the partial disconnection of fresh gas outlet. But there were no alarm signs of anesthetic monitor with adequate ventilator function. This malfunction of anesthetic gas supply without alarm signs may lead to hypoxia and awareness of patients. The intraoperative monitoring of anesthetic gas analyzer and BIS may be essential to detect the malfunction of anesthesia machine.

Keyword

Alarm; Disconnection; Fresh gas outlet; Ventilator

MeSH Terms

Anesthesia
Anesthesia, General*
Anoxia
Humans
Methyl Ethers
Middle Aged
Monitoring, Intraoperative
Nebulizers and Vaporizers
Oxygen
Ventilators, Mechanical
Methyl Ethers
Oxygen
Full Text Links
  • APM
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr