J Korean Radiol Soc.  2008 Feb;58(2):169-175. 10.3348/jkrs.2008.58.2.169.

The Optimal Parameter for Radiation Dose in Pediatric Low Dose Abdominal CT: Cross-sectional Dimensions versus Body Weight

Affiliations
  • 1Department of Radiology and Research Institute of Radiology, Asan Medical Center, University of Ulsan, College of Medicine, Korea. hwgoo@amc.seoul.kr

Abstract

PURPOSE: To investigate the best parameter between cross-sectional dimensions and body weight in pediatric low dose abdominal CT.
MATERIALS AND METHODS
One hundred and thirty six children consecutively underwent weight-based abdominal CT. The subjects consisted of group 1 (79 children, weight range 10.0-19.9 kg) and group 2 (57 children, weight range 20.0-39.9 kg). Abdominal cross-sectional dimensions including circumference, area, anteroposterior diameters and transverse diameters were calculated. Image noise (standard deviation of CT density) was measured by placing a region of interest in the posterior segment of the right hepatic lobe on a CT image at the celiac axis. The measured image noise was correlated with the cross-sectional abdominal dimensions and body weight for subjects in each group.
RESULTS
In group 1 subjects, area, circumference, transverse diameter, anteroposterior diameter, and body weight showed a significant positive correlation with image noise in descending order (gamma = 0.63, 0.62, 0.61, 0.51, and 0.49; p<0.0001). In group 2 subjects, transverse diameter, circumference, area, anteroposterior diameter, and body weight showed a significant positive correlation with image noise in descending order (gamma = 0.83, 0.82, 0.78, 0.71, and 0.71; p<0.0001).
CONCLUSION
Cross-sectional dimensions such as area, circumference, and transverse diameter showed a higher positive correlation with image noise than body weight for pediatric low dose abdominal CT.

Keyword

Children; Abdomen, CT; Computed tomography (CT), radiation exposure; Computed tomography (CT), quality assurance; Body weight

MeSH Terms

Axis, Cervical Vertebra
Body Weight
Child
Humans
Noise

Cited by  1 articles

Is It Better to Enter a Volume CT Dose Index Value before or after Scan Range Adjustment for Radiation Dose Optimization of Pediatric Cardiothoracic CT with Tube Current Modulation?
Hyun Woo Goo
Korean J Radiol. 2018;19(4):692-703.    doi: 10.3348/kjr.2018.19.4.692.

Full Text Links
  • JKRS
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr