J Lab Med Qual Assur.  2014 Dec;36(4):196-204. 10.15263/jlmqa.2014.36.4.196.

Increasing the Efficiency of Laboratory Performance by Using the Onboard Dilution Algorithm of the Elecsys Hepatitis B Surface Antigen II Quantitative Assay

Affiliations
  • 1Department of Laboratory Medicine, National Cancer Center, Goyang, Korea.
  • 2Department of Laboratory Medicine and Genetics, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Korea. eskang@skku.edu
  • 3Department of Laboratory Medicine, Seoul St. Mary's Hospital, The Catholic University of Korea College of Medicine, Seoul, Korea.
  • 4Department of Laboratory Medicine, Severance Hospital, Yonsei University College of Medicine, Seoul, Korea.

Abstract

BACKGROUND
The Elecsys hepatitis B surface antigen (HBsAg) II quantitative assay is a newly introduced electrochemiluminescence immunoassay incorporating an initial 1:400 onboard dilution and a simple algorithm to determine HBsAg levels in serum. We evaluated the performance of the Elecsys HBsAg II assay and determined the impact of its initial onboard dilution on laboratory efficiency.
METHODS
HBsAg levels were determined using both Roche Elecsys and Abbott Architect HBsAg assays. Linearity and precision of the Elecsys HBsAg II assay and its correlation with the Architect HBsAg assay were evaluated. In particular, precision was verified at Samsung Medical Center, Severance Hospital, Seoul St. Mary's Hospital in Seoul, using the same pooled serum controls. The efficiency of the dilution algorithm for both methods was verified using data from 1,848 clinical samples.
RESULTS
The Elecsys HBsAg II assay showed a good linearity from 0.1 to 48,000.0 IU/mL and a good correlation (r=0.9998) between expected and measured values. Precision analyses performed at Samsung Medical Center, Severance Hospital, Seoul St. Mary's Hospital showed excellent performance with coefficients of variation between 1.28% and 6.82%. The values of the Elecsys HBsAg II and Architect HBsAg assays were well correlated (n=506, r=0.987, P<0.001) and also reliably determined in hepatitis C virus- and hepatitis B virus-co-infected patient sera (n=27). In terms of efficiency, 64.0% of samples provided a final HBsAg result on the first run without the need for further dilution, when using the 1:400 onboard pre-dilution protocol of the Elecsys HBsAg II assay.
CONCLUSIONS
Given the excellent precision and correlation with the Architect assay, the Elecsys HBsAg II assay showed a potential advantage for laboratory efficiency by significantly reducing the need for retesting samples with high HBsAg levels.

Keyword

Hepatitis B surface antigens; Elecsys HBsAg II; Architect HBsAg; Hepatitis B virus; Chronic hepatitis B

MeSH Terms

Hepatitis B
Hepatitis B Surface Antigens*
Hepatitis B virus
Hepatitis B, Chronic
Hepatitis C
Humans
Immunoassay
Seoul
Hepatitis B Surface Antigens
Full Text Links
  • JLMQA
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr