J Lab Med Qual Assur.  2018 Jun;40(2):109-111. 10.15263/jlmqa.2018.40.2.109.

Performance Evaluation of Elecsys Vitamin D Total II Assay Using Roche Modular Analytics E170

Affiliations
  • 1Department of Laboratory Medicine, Uijeongbu St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Uijeongbu, Korea. hkl@cathlic.ac.kr

Abstract

We evaluated the performance of a recently developed immunoassay, Elecsys vitamin D total II assay (Roche Diagnostics GmbH, Germany). Precision, linearity, and comparison studies were performed according to the Clinical and Laboratory Standards Institute guidelines. Control materials, linearity materials, and patient samples were used for the evaluation. For the correlation study, liquid chromatography-tandem mass spectrometry was used as the comparative method. Total coefficients of variations of the analyte were between 5.1% and 9.6%. The results of linearity evaluation were also acceptable for the range tested. Correlations with comparative methods were good; however, the mean values of the Elecsys vitamin D assay were higher than those of liquid chromatography-tandem mass spectrometry. The overall analytical performance of the Elecsys vitamin D total II assay is acceptable for the immunology analyzer. Therefore, the Elecsys vitamin D total II assay is expected to be widely used.

Keyword

Vitamin D; Performance; Evaluation

MeSH Terms

Allergy and Immunology
Humans
Immunoassay
Mass Spectrometry
Methods
Statistics as Topic
Vitamin D*
Vitamins*
Vitamin D
Vitamins
Full Text Links
  • JLMQA
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr