J Korean Acad Prosthodont.  2006 Jun;44(3):325-332.

Effect of implant designs on insertion torque and implant stability quotient (ISQ) value

Affiliations
  • 1Department of Prosthodontics, College of Dentistry, Seoul National University, Seoul, Korea
  • 2Department of Prosthodontics, College of Dentistry, Yonsei University, Seoul, Korea
  • 3Department of Stomatology, College of Medicine, Affiliated Hospital of Yanbian University, Yanbian, JiLin, China

Abstract

Statement of problem. Primary implant stability has long been identified as a prerequisite to achieve osseointegration. So the application of a simple, clinically applicable noninvasive test to assess implant stability and osseiointegratation are considered highly desirable. Purpose. The purpose of this study was to evaluate the ISQ value and the insertion torque of the 3 different implant system, then to evaluate whether there was a correlation between ISQ value and insertion torque; and to determine whether implant design has an influence on either insertion torque or ISQ value. Material and method. The experiment was composed of 3 groups: depending on the implant fixture design. Group1 was Bra.nemark type parallel implant in 3.75*7mm. Group2 was Oneplant type straight implant in 4.3*8.5mm. Group3 was Oneplant type tapered implant in 4.3*8.5mm. Depending on the density of the bone, 2 types of bone were used in this experiment. Type I bone represented for cortical bone, type II bone represented for cancellous bone. With the insertion of the implant in type I and type II bone, the insertion torque was measured, then the ISQ value was evaluated, and then the correlation between insertion torque and ISQ value was analyzed Result and conclusion. Within the limitations of this study, the following conclusions were drawn. 1. Within the 3 different implants, the insertion torque value and ISQ value were higher in type I bone, when compared with type II bone.(p<0.05) 2. In type I and type II bone, Oneplant type tapered implant has the highest value in insertion torque.(p<0.05) 3. In type I and type II bone, there was no difference in ISQ values among the 3 types of implant. (p>0.05) 4. Significant linear correlation was found in Bra.nemark type parallel implant: 3.75*7mm in type II bone.

Keyword

Implant stability; Insertion torque; Resonance frequency analysis; Implant design

MeSH Terms

Osseointegration
Torque*
Full Text Links
  • JKAP
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr