Korean J Blood Transfus.  1996 Jun;7(1):49-55.

CS-3000 Plus Cell Separator; A-35 Collection Chamber vs PLT-30TM Collection Chamber

Abstract

BACKGROUND
Recently, it has been established that plateletpheresis needs more efficiency and shorter processing time. Fenwall laboratories developed a new collection chamber for CS-3000 Plus, PLT-30TM collection chamber, which can reduce the processing time with efficient collection. We evaluated the PLT-30TM collection chamber by comparing it with A-35 collection chamber that has been used as a standard collection chamber of CS-3000 Plus us.
METHODS
Thirty platelet collection procedures were performed using the CS 3000 Plus with A-35 collection chamber and PLT-30TM collection chamber. The changes of the hematologic parameters between pre- and post-donation in donors and the total platelets yields and the contaminated WBCs in the plateletpheresis products were evaluated. In processing, the yield predictor calibration was adjusted to 1.00 and 1.13 in A-35 and PLT-30TM respectively. Yield predictors of pheresis were the same as 3.5x1011 in both and end point volumes were calculated from the CS-3000 Plus. Processing volume and processing times were compared between A-35 and PLT-30TM groups.
RESULTS
With PLT-30TM collection chamber, 3.38+/-0.72x1011/L platelets were harvested, whereas 3.20+/- 0.73x1011/L were collected with A-35 collection chamber, which was not significantly different. But processing time with the PLT-30TM collection chamber was more reduced than that with the A-35 collection chamber by about 20 minutes (PLT-30TM : 88.6+/-8.4 min, A-35 : 106.7+/-11.7min). Collection efficiency of PLT-30TM chamber was 50.7+/-12.5% and that of A-35 chamber was 44.4 + 8.8%. The leukocyte contamination of the platelet concentrates were not statistically different(PLT-30TM: 0.0-3.6x106, A-35 : 0.1-4.1x106).
CONCLUSIONS
PLT-30TM collection chamber has the advantages of shortening the donation time and decreasing the processing volume with better collection efficiency and flexibility of platelet concentrate volume.


MeSH Terms

Blood Component Removal
Blood Platelets
Calibration
Humans
Leukocytes
Plateletpheresis
Pliability