Korean Circ J.  1998 Jan;28(1):89-96. 10.4070/kcj.1998.28.1.89.

Comparison of Porcine Corinary Stent Restenosis between MAC (Maximum Arterial Re-Creation)and Palmaz-Schatz Stent

Abstract

BACKGROUND
Coronary stents are effective in the treatment of acute complications after angioplasty and the prevention of restenosis. However, complications may arise posttreatment, the major cinical problems are stent thrombosis and restenosis. All coronary stents are imported from western countries and are not covered by medical insurance in Korea. Korean stents should be developed to reduce patient's economic burden.
METHODS
We placed five Maximum Arterial Re-Creation (MAC) and five Palmaz-Schatz (PS) stents in ten porcine coronary arteries. Stent and artery diameter ratio was 1.3 : 1.0. Follow-up coronary angiogram and histopathologic examinations were performed four weeks after stent overdilation injury.
RESULTS
All of the stented arteries were patent on follow-up coronary angiograms. Angiographic diameter stenosis was 23.1+/-9.2% in MAC stents and 18.5+/-12.3% in PS stents and pathologic area stenosis was 35.6+/-11.4% in MAC stent and 39.8+/-9.9% in PS stent at 4 weeks after stenting, which were not different between two stents. Maximal intimal thickness (0.28+/-0.12 vs. 0.36+/-0.17mm) and neointimal area (1.83+/-1.01mm2 vs. 1.50+/-0.65mm2) were not different between MAC and PS stent.
CONCLUSIONS
MAC stents are as effective as PS stents in the prevention of stent restenosis in a porcine stent restenosis model.

Keyword

Stent; Restenosis; MAC (Maximum Arterial Re-Creation)Stent

MeSH Terms

Angioplasty
Arteries
Constriction, Pathologic
Coronary Vessels
Follow-Up Studies
Insurance
Korea
Stents*
Thrombosis