Korean Circ J.  1999 May;29(5):498-506. 10.4070/kcj.1999.29.5.498.

The Effects of the Heparin-Coated Maximum Arterial Re-Creation (MAC) Stent on Porcine Coronary Stent Restenosis

Abstract

BACKGROUND
Stent thrombosis and late restenosis are still major limitations in the clinical use of coronary stenting. Heparin-coated stent may reduce the incidences of stent thrombosis and restenosis. Heparin-coated stents were compared with control stents in a porcine coronary stent restenosis model in order to evaluate the effects of heparin-coated stent on stent restenosis.
METHODS
Heparin was coated on a stent by deposition of an ultra-thin polymeric film containing amine groups by means of plasma polymerization. And then stent was immersed in heparin solution. Stent overdilation injury (stent:artery=1.3:1.0) was performed with bare (Group I, n=4) and heparin-coated (Group II, n=5) MAC stents in porcine coronary arteries. Follow-up quantitative coronary angiography (QCA) was performed at 4 weeks after stenting. The histopathologic assessments (KERN=*)of stented porcine coronary arteries were compared in between 2 groups.
RESULTS
1)Luminal area of stented artery was 7.05+/-1.25 mm2 in Group I and 7.67+/-2.85 mm2 in Group II, which were not different between two Groups. 2)Histopathologic stenosis of Group I was 35.7+/-13.2%, which was higher than 28.6+/-14.7% of Group II (p<0.05). Ratio of neointima/media was 1.16+/-0.52 in Group I and 0.87+/-0.31 in Group II and neointimal area was higher in Group I than in Group II (3.81+/-1.78 mm2 vs. 2.82+/-1.11 mm2, p<0.05 respectively). 3)PCNA (Proliferating cell nuclear antigen) index of GroupI was 10.0+/-2.2%, which was higher than in Group II (6.8+/-4.0%).
CONCLUSIONS
Heparin-coated MAC stent may be effective in the inhibition of neointimal proliferation in a porcine stent restenosis model.

Keyword

Heparin-coated stent; Stent Restenosis

MeSH Terms

Arteries
Constriction, Pathologic
Coronary Angiography
Coronary Vessels
Follow-Up Studies
Heparin
Incidence
Plasma
Polymerization
Polymers
Stents*
Thrombosis
Heparin
Polymers

Cited by  1 articles

Inhibitory Effect of Double Coating with Echinomycin and Hydrophobic Heparin in a Porcine Coronary In-Stent Restenosis Model
Sang Yup Lim, Myung Ho Jeong, Doo Sun Sim, Young Joon Hong, Ju Han Kim, Youngkeun Ahn, Jeong Gwan Cho, Jong Chun Park, Jung Chaee Kang, Hyun Tae Moon, Youngro Byun
Chonnam Med J. 2009;45(2):87-91.    doi: 10.4068/cmj.2009.45.2.87.

Full Text Links
  • KCJ
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr