J Korean Orthop Assoc.  2007 Apr;42(2):227-235. 10.4055/jkoa.2007.42.2.227.

Comparison of Radiologic Measurements of Total Knee Replacement using Computer-assisted Navigation System and Conventional System in Varus Deformity of the Knee

Affiliations
  • 1Department of Orthopaedic Surgery, School of Medicine, Kyung Hee University, Seoul, Korea. bdkyung@khmc.or.kr
  • 2Department of Orthopaedic Surgery, Sejong Hospital, Bucheon, Korea.

Abstract

PURPOSE: To compare the radiologic measurements of the mechanical axis and the implant position of Total Knee Arthroplasty (TKA) using a computer-assisted navigation system with those using conventional TKA in varus deformity. MATERIALS AND METHODS: From January 2004 to January 2005, 49 TKAs using a CT-free navigation system (Vector Vision(R), BrainLab, Heirnstetten, Germany) (Group I) and 24 TKAs using the conventional technique (Group II) were performed on patients who had a preoperative varus deformity>10degrees. The patients were also subdivided into two groups, patients with a varus deformity<20degrees (group A) and patients with varus deformity > 20degrees (group B). The PFC Sigma implants were used in both groups. The mechanical axis and implant position were measured by 2 observers according to the reontgenographic evaluation system of the American Knee Society. RESULTS: There was no significant difference in alpha, beta, delta angle and mechanical axis between group I and II. There was a significant difference in the gamma angle between group I and II (p<0.05). There was a significant difference in the alpha and beta angle and mechanical axis between group IA and IB (p<0.05). There was a significant difference in the alpha angle and mechanical axis between group IIA and IIB (p<0.05). There was a positive correlation between the measured angle by the respective observers in all groups (p<0.05). CONCLUSION: Patients with a preoperative varus deformity>20degrees tended to have more postoperative varus mechanical alignment than those with a preoperative varus deformity between 10degrees and 20degrees after TKA. More careful attention during the registration of the femoral mechanical axis should be paid in patients with a larger varus deformity in TKA using a computer-assisted navigation system. On the other hand, a reasonable mechanical valgus angle should be considered in femoral bone cutting for a varus deformity of the distal femur in conventional TKA. In addition, inadequate positioning of intramedullary rod should be recognized in conventional TKA.

Keyword

Knee; Varus deformity; Arthroplasty; Computer-assisted navigation system; Radiologic measurements

Figure

  • Fig. 1 (A) The femoral component angle (A) was measured between a line parallel to the femoral condyle and the mechanical axis of the femur on the anteroposterior view. (B) The tibial mechanical angle (B) was measured between a line parallel to the tibial condyle (a) and a line parallel to the ankle joint (b).

  • Fig. 2 Post-operative measurement of the femoral and tibial component. α, The inclination angle of the femoral component was formed with the anatomical axis of the femur. β, The inclination angle of the tibial component was formed with the mechanical axis of the tibial. γ, The angle of the femoral component was formed between the anatomical axis of the femur and perpendicular line to the prosthesis. δ, The angle of the tibial component was formed with the mechanical axis of the tibial.

  • Fig. 3 Comparison of the different distributions of the γ angle between Group I and Group II.

  • Fig. 4 Comparison of the different distributions of the postoperative mechanical axis between Group IA and IB.

  • Fig. 5 Comparison of the different distributions of the postoperative mechanical axis between Group IIA and IIB (ed note: I think the highlighted can be deleted).


Cited by  1 articles

Computer Assisted Navigation in Knee Arthroplasty
Dae Kyung Bae, Sang Jun Song
Clin Orthop Surg. 2011;3(4):259-267.    doi: 10.4055/cios.2011.3.4.259.


Reference

1. Cates HE, Ritter MA, Keating EM, Faris PM. Intramedullary versus extramedullary femoral alignment systems in total knee replacement. Clin Orthop Relat Res. 1993. 286:32–39.
Article
2. Engh GA, Petersen TL. Comparative experience with intramedullary and extramedullary alignment in total knee arthroplasty. J Arthroplasty. 1990. 5:1–8.
Article
3. Garg A, Walker PS. Prediction of total knee motion using a three-dimensional computer-graphics model. J Biomech. 1990. 23:45–58.
Article
4. Ishii Y, Ohmori G, Bechtold JE, Gustilo RB. Extramedullary versus intramedullary alignment guides in total knee arthroplasty. Clin Orthop Relat Res. 1995. 318:167–175.
5. Jiang CC, Insall JN. Effect of rotation on the axial alignment of the femur. Pitfalls in the use of femoral intramedullary guides in total knee arthroplasty. Clin Orthop Relat Res. 1989. 248:50–56.
6. Laskin RS. Instrumentation pitfalls: you just can't go on autopilot! J Arthroplasty. 2003. 18:18–22.
Article
7. Lotke PA, Ecker ML. Influence of positioning of prosthesis in total knee replacement. J Bone Joint Surg Am. 1977. 59:77–79.
Article
8. Mahaluxmivala J, Bankes MJ, Nicolai P, Aldam CH, Allen PW. The effect of surgeon experience on component positioning in 673 Press Fit Condylar posterior cruciate-sacrificing total knee arthroplasties. J Arthroplasty. 2001. 16:635–640.
Article
9. Mielke RK, Clemens U, Jens JH, Kershally S. Navigation in knee endoprosthesis implantation--preliminary experiences and prospective comparative study with conventional implantation technique. Z Orthop Ihre Grenzgeb. 2001. 139:109–116.
10. Mihalko WM, Boyle J, Clark LD, Krackow KA. The variability of intramedullary alignment of the femoral component during total knee arthroplasty. J Arthroplasty. 2005. 20:25–28.
Article
11. Moreland JR. Mechanisms of failure in total knee arthroplasty. Clin Orthop Relat Res. 1988. 226:49–64.
Article
12. Oswald MH, Jakob RP, Schneider E, Hoogewoud HM. Radiological analysis of normal axial alignment of femur and tibia in view of total knee arthroplasty. J Arthroplasty. 1993. 8:419–426.
Article
13. Reed SC, Gollish J. The accuracy of femoral intramedullary guides in total knee arthroplasty. J Arthroplasty. 1997. 12:677–682.
Article
14. Ritter MA, Faris PM, Keating EM, Meding JB. Postoperative alignment of total knee replacement. Its effect on survival. Clin Orthop Relat Res. 1994. 299:153–156.
Article
15. Sambatakakis A, Wilton TJ, Newton G. Radiographic sign of persistent soft-tissue imbalance after knee replacement. J Bone Joint Surg Br. 1991. 73:751–756.
Article
16. Seon JK, Song EK. The accuracy of lower extremity alignment in a total knee arthroplasty using computer-assisted navigation system. J Korean Orthop Assoc. 2004. 39:566–571.
Article
17. Sparmann M, Wolke B, Czupalla H, Banzer D, Zink A. Positioning of total knee arthroplasty with and without navigation support. A prospective, randomised study. J Bone Joint Surg Br. 2003. 85:830–835.
18. Stulberg SD, Loan P, Sarin V. Computer-assisted navigation in total knee replacement: results of an initial experience in thirty-five patients. J Bone Joint Surg Am. 2002. 84:Suppl 2. S90–S98.
19. Tang WM, Chiu KY, Kwan MF, Ng TP, Yau WP. Sagittal bowing of the distal femur in Chinese patients who require total knee arthroplasty. J Orthop Res. 2005. 23:41–45.
Article
Full Text Links
  • JKOA
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles