J Adv Prosthodont.  2024 Aug;16(4):221-230. 10.4047/jap.2024.16.4.221.

Recording maximal intercuspation and border positions of the mandible with intraoral scanner using the acquisition software’s multi-occlusion function

Affiliations
  • 1Fixed Prosthodontics, Department of Conservative Dentistry, Alexandria University, Alexandria, Egypt

Abstract

PURPOSE
This in vitro study was conducted to investigate the accuracy of intraoral scanner (IOS) for recording maximal intercuspal position (MIP) and border positions of the mandible.
MATERIALS AND METHODS
Maxillary and mandibular master casts were articulated in MIP, protrusive, and lateral interocclusal position sequentially on a semi-adjustable articulator. For each articulation relation, sites of occlusal contacts (SOCs) and sites of clearance (SCs) were identified on the master casts with articulating paper (reference sites). IOS was used to take full arch scans and nine virtual interocclusal records (VIRs) for virtual articulation of models. Virtual SOCs and SCs were detected with 3D processing software and compared to those identified with the articulating paper. Sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) were calculated for each articulation relation.
RESULTS
For MIP, IOS showed adequate sensitivity and NPV of 100%, and specificity and a PPV of 99%. For protrusive position, the IOS showed a sensitivity and a NPV of 100%, a high PPV of 86%, and a specificity of 83%. For lateral positions, the specificity and the PPV were high (93% and 79%, respectively), but the sensitivity and the NPV were below the clinically acceptable limits (28% and 56%, respectively).
CONCLUSION
IOS displayed clinically acceptable accuracy for recording MIP and protrusive border mandibular position. However, IOS had less accuracy for lateral border mandibular position.

Keyword

Intraoral scanner; Mandibular position; Maximal intercuspal position; Occlusion; Virtual