J Korean Med Sci.  2024 Jan;39(2):e18. 10.3346/jkms.2024.39.e18.

First-in-Human Study for Evaluating the Accuracy of Smart Ring Based Cuffless Blood Pressure Measurement

Affiliations
  • 1Division of Cardiology, Heart Vascular Stroke Institute, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Korea

Abstract

Background
Recently, a ring-type cuffless blood pressure (BP) measuring device has been developed. This study was a prospective, single arm, first-in-human pivotal trial to evaluate accuracy of BP measurement by the new device.
Methods
The ring-type smart wearable monitoring device measures photoplethysmography signals from the proximal phalanx and transmits the data wirelessly to a connected smartphone. For the BP comparison, a cuff was worn on the arm to check the reference BP by auscultatory method, while the test device was worn on the finger of the opposite arm to measure BP simultaneously. Measurements were repeated for up to three sets each on the left and right arms. The primary outcome measure was mean difference and standard deviation of BP differences between the test device and the reference readings.
Results
We obtained 526 sets of systolic BP (SBP) and 513 sets of diastolic BP (DBP) from 89 subjects, with ranges of 80 to 175 mmHg and 43 to 122 mmHg for SBP and DBP, respectively. In sample-wise comparison, the mean difference between the test device and the reference was 0.16 ± 5.90 mmHg (95% limits of agreement [LOA], −11.41, 11.72) in SBP and −0.07 ± 4.68 (95% LOA, −9.26, 9.10) in DBP. The test device showed a strong correlation with the reference for SBP (r = 0.94, P < 0.001) and DBP (r = 0.95, P < 0.001). There were consistent results in subject-wise comparison.
Conclusion
The new ring-type BP measuring device showed a good correlation for SBP and DBP with minimal bias compared with an auscultatory method.

Keyword

Blood Pressure Monitors; Photoplethysmography; Wearable Electronic Devices

Figure

  • Fig. 1 The ring-type BP measuring device. The ring-type smart wearable BP monitoring device measures and analyzes PPG signals from the proximal phalanx and transmits the data wirelessly to a connected smartphone.LED = light emitting diode, PD = photo diode, BP = blood pressure, PPG = photoplethysmography, SYS = systolic, DIA = diastolic.

  • Fig. 2 Study flow.BP = blood pressure, SBP = systolic blood pressure, DBP = diastolic blood pressure, FA = full analysis, PP = per-protocol.aThese criteria are based on the International Organization for Standardization 81060-2.

  • Fig. 3 Correlation of blood pressure measures between the ring-type device and auscultatory method. (A) Correlation for SBP between the test device and reference, (B) correlation for DBP between the test device and reference, (C) Bland-Altman plot for SBP between the methods, (D) Bland-Altman plot for DBP between the methods.SBP = systolic blood pressure, DBP = diastolic blood pressure, LOA = limits of agreement, SD = standard deviation.


Reference

1. NCD Risk Factor Collaboration (NCD-RisC). Worldwide trends in hypertension prevalence and progress in treatment and control from 1990 to 2019: a pooled analysis of 1201 population-representative studies with 104 million participants. Lancet. 2021; 398(10304):957–980. PMID: 34450083.
2. US Preventive Services Task Force. Krist AH, Davidson KW, Mangione CM, Cabana M, Caughey AB, et al. Screening for hypertension in adults: US preventive services task force reaffirmation recommendation statement. JAMA. 2021; 325(16):1650–1656. PMID: 33904861.
3. GBD 2017 Risk Factor Collaborators. Global, regional, and national comparative risk assessment of 84 behavioural, environmental and occupational, and metabolic risks or clusters of risks for 195 countries and territories, 1990-2017: a systematic analysis for the Global Burden of Disease Study 2017. Lancet. 2018; 392(10159):1923–1994. PMID: 30496105.
4. Williams B, Mancia G, Spiering W, Agabiti Rosei E, Azizi M, Burnier M, et al. 2018 ESC/ESH guidelines for the management of arterial hypertension. Eur Heart J. 2018; 39(33):3021–3104. PMID: 30165516.
5. Unger T, Borghi C, Charchar F, Khan NA, Poulter NR, Prabhakaran D, et al. 2020 International Society of Hypertension global hypertension practice guidelines. Hypertension. 2020; 75(6):1334–1357. PMID: 32370572.
6. Padwal R, Campbell NR, Schutte AE, Olsen MH, Delles C, Etyang A, et al. Optimizing observer performance of clinic blood pressure measurement: a position statement from the Lancet Commission on Hypertension Group. J Hypertens. 2019; 37(9):1737–1745. PMID: 31034450.
7. Clement DL, De Buyzere ML, De Bacquer DA, de Leeuw PW, Duprez DA, Fagard RH, et al. Prognostic value of ambulatory blood-pressure recordings in patients with treated hypertension. N Engl J Med. 2003; 348(24):2407–2415. PMID: 12802026.
8. Dolan E, Stanton A, Thijs L, Hinedi K, Atkins N, McClory S, et al. Superiority of ambulatory over clinic blood pressure measurement in predicting mortality: the Dublin outcome study. Hypertension. 2005; 46(1):156–161. PMID: 15939805.
9. Niiranen TJ, Mäki J, Puukka P, Karanko H, Jula AM. Office, home, and ambulatory blood pressures as predictors of cardiovascular risk. Hypertension. 2014; 64(2):281–286. PMID: 24842916.
10. Oh YK, Chin HJ, Ahn SY, An JN, Lee JP, Lim CS, et al. Discrepancies in clinic and ambulatory blood pressure in Korean chronic kidney disease patients. J Korean Med Sci. 2017; 32(5):772–781. PMID: 28378550.
11. Viera AJ, Lingley K, Hinderliter AL. Tolerability of the Oscar 2 ambulatory blood pressure monitor among research participants: a cross-sectional repeated measures study. BMC Med Res Methodol. 2011; 11(1):59. PMID: 21524301.
12. Huang QF, Yang WY, Asayama K, Zhang ZY, Thijs L, Li Y, et al. Ambulatory blood pressure monitoring to diagnose and manage hypertension. Hypertension. 2021; 77(2):254–264. PMID: 33390042.
13. Castaneda D, Esparza A, Ghamari M, Soltanpur C, Nazeran H. A review on wearable photoplethysmography sensors and their potential future applications in health care. Int J Biosens Bioelectron. 2018; 4(4):195–202. PMID: 30906922.
14. Joung J, Jung CW, Lee HC, Chae MJ, Kim HS, Park J, et al. Continuous cuffless blood pressure monitoring using photoplethysmography-based PPG2BP-net for high intrasubject blood pressure variations. Sci Rep. 2023; 13(1):8605. PMID: 37244974.
15. Liu C, Zheng D, Griffiths C, Murray A. Comparison of repeatability of blood pressure measurements between oscillometric and auscultatory methods. Comput Cardiol. 2015; 42:1073–1076.
16. Lewis PS. British and Irish Hypertension Society’s Blood Pressure Measurement Working Party. Oscillometric measurement of blood pressure: a simplified explanation. A technical note on behalf of the British and Irish Hypertension Society. J Hum Hypertens. 2019; 33(5):349–351. PMID: 30926901.
17. Allen J. Photoplethysmography and its application in clinical physiological measurement. Physiol Meas. 2007; 28(3):R1–39. PMID: 17322588.
18. Slapničar G, Mlakar N, Luštrek M. Blood pressure estimation from photoplethysmogram using a spectro-temporal deep neural network. Sensors (Basel). 2019; 19(15):3420. PMID: 31382703.
19. Hosanee M, Chan G, Welykholowa K, Cooper R, Kyriacou PA, Zheng D, et al. Cuffless single-site photoplethysmography for blood pressure monitoring. J Clin Med. 2020; 9(3):723. PMID: 32155976.
20. Elgendi M, Fletcher R, Liang Y, Howard N, Lovell NH, Abbott D, et al. The use of photoplethysmography for assessing hypertension. NPJ Digit Med. 2019; 2:60. PMID: 31388564.
21. Picone DS, Deshpande RA, Schultz MG, Fonseca R, Campbell NR, Delles C, et al. Nonvalidated home blood pressure devices dominate the online marketplace in Australia: major implications for cardiovascular risk management. Hypertension. 2020; 75(6):1593–1599. PMID: 32275193.
22. Wang L, Xian H, Guo J, Li W, Wang J, Chen Q, et al. A novel blood pressure monitoring technique by smart HUAWEI WATCH: a validation study according to the ANSI/AAMI/ISO 81060-2:2018 guidelines. Front Cardiovasc Med. 2022; 9:923655. PMID: 36304535.
23. Schoettker P, Degott J, Hofmann G, Proença M, Bonnier G, Lemkaddem A, et al. Blood pressure measurements with the OptiBP smartphone app validated against reference auscultatory measurements. Sci Rep. 2020; 10(1):17827. PMID: 33082436.
24. Armstrong MK, Schultz MG, Picone DS, Black JA, Dwyer N, Roberts-Thomson P, et al. Brachial and radial systolic blood pressure are not the same. Hypertension. 2019; 73(5):1036–1041. PMID: 30905194.
25. Picone DS, Schultz MG, Otahal P, Aakhus S, Al-Jumaily AM, Black JA, et al. Accuracy of cuff-measured blood pressure: systematic reviews and meta-analyses. J Am Coll Cardiol. 2017; 70(5):572–586. PMID: 28750701.