Clin Orthop Surg.  2021 Dec;13(4):449-455. 10.4055/cios20202.

Clinical Results of Dual SC Screw: A Mini-Sliding Hip Screw with an Anti-rotating Screw for Femoral Neck Fractures

Affiliations
  • 1Department of Orthopaedic Surgery and Joint Surgery Centre, Takatsuki General Hospital, Takatsuki, Japan

Abstract

Background
Dual SC screw (DSCS) is a unique concept internal fixation device consisting of a sliding screw and barrel assembly that enables compression force to be applied to the femoral neck fracture side. There are two types of barrels: a thread barrel and a plate barrel that has a one-holed side plate to prevent varus deformity. We report clinical results of the application of a DSCS with combined use of a thread barrel screw as an anti-rotational screw and a plate barrel screw as a compression hip screw.
Methods
We used DSCS for femoral neck fractures in 196 hip joints of 190 patients between November 2005 and June 2017. Among them, 70 hips in 66 patients (13 men and 53 women; mean age, 73.2 years) were followed up for at least 24 months. There were 53 nondisplaced fractures (Garden’s classification stage 1 or 2) and 17 displaced fractures (stage 3 or 4). We evaluated the postoperative walking ability of the patients who were followed up for at least 24 months and examined details of all complications.
Results
The mean follow-up period was 37.4 months (range, 24–144 months). Forty-two (64%) out of the 66 patients who were followed up for at least 2 years were able to walk independently with or without a cane. The incidence of complications was 11.5% in nondisplaced fractures and 17.5% in displaced fractures, and arthroplasty was required in 5.8% and 5.0%, respectively. The most frequent complication was secondary displacement including cutout and shortening of the femoral neck, but no implants showed varus displacement.
Conclusions
The application of DSCS for the treatment of femoral neck fractures had satisfactory results. The complication rate was low, and there was no postoperative varus displacement of DSCS in either displaced or nondisplaced fractures. We suggest DSCS is a reliable option for both displaced and nondisplaced femoral neck fractures.

Keyword

Femoral neck fracture; Treatment; Operation; Screw; Complication
Full Text Links
  • CIOS
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr