Korean J Radiol.  2021 Oct;22(10):1708-1718. 10.3348/kjr.2020.1455.

Investigation of the Characteristics of New, Uniform, Extremely Small Iron-Based Nanoparticles as T1 Contrast Agents for MRI

Affiliations
  • 1Department of Radiology, SMG-SNU Boramae Medical Center, Seoul, Korea
  • 2Department of Radiology, Seoul National University College of Medicine, Seoul, Korea
  • 3Department of Radiology, Seoul National University Hospital, Seoul, Korea
  • 4Institute of Radiation Medicine, Seoul National University Medical Research Center, Seoul, Korea
  • 5Department of Radiology and Research Institute of Radiological Science, Severance Hospital, Yonsei University College of Medicine, Seoul, Korea

Abstract


Objective
The purpose of this study was to evaluate the magnetic resonance (MR) characteristics and applicability of new, uniform, extremely small iron-based nanoparticles (ESIONs) with 3–4-nm iron cores using contrast-enhanced magnetic resonance angiography (MRA).
Materials and Methods
Seven types of ESIONs were used in phantom and animal experiments with 1.5T, 3T, and 4.7T scanners. The MR characteristics of the ESIONs were evaluated via phantom experiments. With the ESIONs selected by the phantom experiments, animal experiments were performed on eight rabbits. In the animal experiments, the in vivo kinetics and enhancement effect of the ESIONs were evaluated using half-diluted and non-diluted ESIONs. The between-group differences were assessed using a linear mixed model. A commercially available gadolinium-based contrast agent (GBCA) was used as a control.
Results
All ESIONs showed a good T1 shortening effect and were applicable for MRA at 1.5T and 3T. The relaxivity ratio of the ESIONs increased with increasing magnetic field strength. In the animal experiments, the ESIONs showed peak signal intensity on the first-pass images and persistent vascular enhancement until 90 minutes. On the 1-week follow-up images, the ESIONs were nearly washed out from the vascular structures and organs. The peak signal intensity on the first-pass images showed no significant difference between the non-diluted ESIONs with 3-mm iron cores and GBCA (p = 1.000). On the 10-minutes post-contrast images, the non-diluted ESIONs showed a significantly higher signal intensity than did the GBCA (p < 0.001).
Conclusion
In the phantom experiments, the ESIONs with 3–4-nm iron oxide cores showed a good T1 shortening effect at 1.5T and 3T. In the animal experiments, the ESIONs with 3-nm iron cores showed comparable enhancement on the first-pass images and superior enhancement effect on the delayed images compared to the commercially available GBCA at 3T.

Keyword

Iron oxide nanoparticle; Blood pool contrast agent; Magnetic resonance angiography; Contrast enhancement