Utility and Diagnostic Performance of Automated Breast Ultrasound System in Evaluating Pure Non-Mass Enhancement on Breast Magnetic Resonance Imaging
- Affiliations
-
- 1Department of Radiology, Seoul National University Hospital, Seoul National University College of Medicine, Seoul, Korea.
- 2Department of Radiology, Seoul National University Hospital Healthcare System Gangnam Center, Seoul, Korea.
Abstract
Objective
To compare the utility and diagnostic performance of automated breast ultrasound system (ABUS) with that of handheld ultrasound (HHUS) in evaluating pure non-mass enhancement (NME) lesions on breast magnetic resonance imaging (MRI).
Materials and Methods
One hundred twenty-six consecutive MRI-visible pure NME lesions of 122 patients with breast cancer were assessed from April 2016 to March 2017. Two radiologists reviewed the preoperative breast MRI, ABUS, and HHUS images along with mammography (MG) findings. The NME correlation rate and diagnostic performance of ABUS were compared with that of HHUS, and the imaging features associated with ABUS visibility were analyzed.
Results
Among 126 pure NME lesions, 100 (79.4%) were malignant and 26 (20.6%) were benign. The overall correlation rate was 87.3% (110/126) in ABUS and 92.9% (117/126) in HHUS. The sensitivity and specificity were 87% and 50% for ABUS and 92% and 42.3% for HHUS, respectively, with no significant differences (p = 0.180 and 0.727, respectively). Malignant NME was more frequently visualized than benign NME lesions on ABUS (93% vs. 65.4%, p = 0.001). Significant factors associated with the visibility of ABUS were the size of NME lesions on MRI (p < 0.001), their distribution pattern (p < 0.001), and microcalcifications on MG (p = 0.027).
Conclusion
ABUS evaluation of pure NME lesions on MRI in patients with breast cancer is a useful technique with high visibility, especially in malignant lesions. The diagnostic performance of ABUS was comparable with that of conventional HHUS in evaluating NME lesions.