J Lab Med Qual Assur.  2019 Mar;41(1):29-38. 10.15263/jlmqa.2019.41.1.29.

The Performance of the Agility System for Interferon Gamma Release Assay Using QuantiFERON TB Gold In-Tube Assay

Affiliations
  • 1Laboratory Medicine Center, The Korean Institute of Tuberculosis, Cheongju, Korea. leukoso@naver.com

Abstract

BACKGROUND
As stated in "˜The Action Strategy for Tuberculosis-Free Korea,' last March, high-throughput, large-scale analytical instruments for interferon gamma release assays (IGRA) are demanded by many clinical laboratories using the QuantiFERON-TB Gold In-Tube assay (Cellestis/Qiagen, Australia). Agility (Dynex Technologies, USA) is an automated high-throughput enzyme linked immunosorbent assay analyser. The present study aimed to evaluate its accuracy and speed.
METHODS
Pooled plasma was prepared using samples obtained after IGRA testing. Analyses of precision, linearity, cut-off evaluation, and comparison with conventional methods were performed for multiple Agility instruments according to the Clinical and Laboratory Standards Institute EP5-A3, EP6-A, EP9-A3 and EP12-A2 guidelines. The turnaround time and throughput were also analysed.
RESULTS
The coefficient of variation range was 2.48%-4.0%, 7.01%-11.17%, and 9.69%-14.84% for the repeatability, between-run precision, and between-day precision analyses, respectively. The linearity ranged from 0 to 10.541. Comparison analysis presented a high concordance of Agility with the conventional instrument, DS2 (Dynex Technologies), and manual method for IGRA. The cut-off value of 0.35 IU/mL was well compatible with the C50. It was identified that the C50±20% contained the C5-C95 interval. The average turnaround time was 3.84 hours, from the submission of pre-treated samples to the reporting of results. The throughput was determined to be 290 tests during a routine working time of 8 hours.
CONCLUSIONS
Agility showed high precision, linearity, concordance, and had a 2.5 times faster throughput than with the conventional and manual method. It could be useful for large-scale IGRA testing in latent tuberculosis infection screening project. Samples within C50±20% are suspected to show relatively low reporducible results of high inversion between postivie and negative.

Keyword

Agility; Interferon gamma release assays; Latent tuberculosis infection; QuantiFERON-TB Gold In-Tube

MeSH Terms

Enzyme-Linked Immunosorbent Assay
Interferon-gamma Release Tests
Interferons*
Latent Tuberculosis
Mass Screening
Methods
Plasma
Interferons