Ann Clin Microbiol.  2018 Mar;21(1):12-19. 10.5145/ACM.2018.21.1.12.

Clinical Evaluation of QMAC-dRAST for Direct and Rapid Antimicrobial Susceptibility Test with Gram-Positive Cocci from Positive Blood Culture Bottles

Affiliations
  • 1QuantaMatrix Inc., Seoul, Korea. skwon@snu.ac.kr
  • 2Institutes of Entrepreneurial BioConvergence, Seoul National University, Seoul, Korea.
  • 3Department of Electrical and Computer Engineering, Seoul National University, Seoul, Korea.
  • 4Department of Transdisciplinary Studies, Seoul National University, Seoul, Korea.
  • 5Department of Laboratory Medicine, Seoul National University Hospital, Seoul, Korea.
  • 6Seoul National University Hospital Biomedical Research Institute, Seoul National University Hospital, Seoul, Korea.

Abstract

BACKGROUND
Timely intervention in the treatment of bloodstream infection is important for prescription of appropriate antimicrobials. With prompt determination of the antimicrobial susceptibility of a causative agent, rapid antimicrobial susceptibility test (AST) can help select the appropriate antimicrobial therapy. This clinical study is for evaluation of the clinical performance of the QMAC-dRAST for rapid AST directly from positive blood culture (PBC)s with Gram-positive cocci.
METHODS
A total of 115 PBC samples with Gram-positive organisms (76 Staphylococcus spp. and 39 Enterococcus spp.) were evaluated by the QMAC-dRAST system, and their pure culture isolates were evaluated by the MicroScan WalkAway (Beckman Coulter, USA) as the comparative AST system. Thirteen antimicrobial agents were included, and the agreement and discrepancy rates of the QMAC-dRAST system (Quantamatrix Inc., Republic of Korea) compared to the MicroScan WalkAway were calculated. To resolve discrepancies, the broth microdilution method was performed.
RESULTS
The QMAC-dRAST system exhibited a categorical agreement rate of 94.9% (1,126/1,187) and an essential agreement rate of 98.3% (1,167/1,187). The QMAC-dRAST system yielded very major (false-susceptible) errors at 1.0% (5/485), major (false-resistant) errors at 1.3% (9/693), and minor errors at 4.0% (47/1,187) compared to the MicroScan WalkAway. The QMAC-dRAST system significantly eliminated 30 hours of total turnaround time by combination of direct inoculation of PBC and an image-based approach.
CONCLUSION
The results of the QMAC-dRAST system were highly accurate. Thereby, the QMAC-dRAST may provide essential information to accelerate therapeutic decisions for earlier and adequate antibiotic treatment and patient management in clinical settings.

Keyword

Antimicrobial drug resistance; Bacteremia; Bioengineering; Microbial sensitivity test

MeSH Terms

Anti-Infective Agents
Bacteremia
Bioengineering
Clinical Study
Drug Resistance, Microbial
Enterococcus
Gram-Positive Cocci*
Humans
Methods
Microbial Sensitivity Tests
Prescriptions
Staphylococcus
Anti-Infective Agents

Reference

1. Schefold JC, Hasper D, Volk HD, Reinke P. Sepsis: time has come to focus on the later stages. Med Hypotheses. 2008; 71:203–208.
Article
2. Gherardi G, Angeletti S, Panitti M, Pompilio A, Di Bonaventura G, Crea F, et al. Comparative evaluation of the Vitek-2 Compact and Phoenix systems for rapid identification and antibiotic susceptibility testing directly from blood cultures of Gram-negative and Gram-positive isolates. Diagn Microbiol Infect Dis. 2012; 72:20–31.
Article
3. Barenfanger J, Drake C, Kacich G. Clinical and financial benefits of rapid bacterial identification and antimicrobial susceptibility testing. J Clin Microbiol. 1999; 37:1415–1418.
Article
4. Munoz-Dávila MJ, Yagüe G, Albert M, García-Lucas T. Comparative evaluation of Vitek 2 identification and susceptibility testing of Gram-negative rods directly and isolated from BacT/ALERT-positive blood culture bottles. Eur J Clin Microbiol Infect Dis. 2012; 31:663–669.
Article
5. Bazzi AM, Rabaan AA, Fawarah MM, Al-Tawfiq JA. Direct identification and susceptibility testing of positive blood cultures using high speed cold centrifugation and Vitek II system. J Infect Public Health. 2017; 10:299–307.
Article
6. Jo SJ, Park KG, Han K, Park DJ, Park YJ. Direct identification and antimicrobial susceptibility testing of bacteria from positive blood culture bottles by matrix-assisted laser desorption/ionization time-of-flight mass spectrometry and the Vitek 2 system. Ann Lab Med. 2016; 36:117–123.
Article
7. Pulido MR, García-Quintanilla M, Martín-Peña R, Cisneros JM, McConnell MJ. Progress on the development of rapid methods for antimicrobial susceptibility testing. J Antimicrob Chemother. 2013; 68:2710–2717.
Article
8. Ling TK, Liu ZK, Cheng AF. Evaluation of the VITEK 2 system for rapid direct identification and susceptibility testing of gram-negative bacilli from positive blood cultures. J Clin Microbiol. 2003; 41:4705–4707.
Article
9. de Cueto M, Ceballos E, Martinez-Martinez L, Perea EJ, Pascual A. Use of positive blood cultures for direct identification and susceptibility testing with the Vitek 2 system. J Clin Microbiol. 2004; 42:3734–3738.
Article
10. Ligozzi M, Bernini C, Bonora MG, De Fatima M, Zuliani J, Fontana R. Evaluation of the VITEK 2 system for identification and antimicrobial susceptibility testing of medically relevant gram-positive cocci. J Clin Microbiol. 2002; 40:1681–1686.
Article
11. Funke G, Funke-Kissling P. Use of the BD PHOENIX Automated Microbiology System for direct identification and susceptibility testing of gram-negative rods from positive blood cultures in a three-phase trial. J Clin Microbiol. 2004; 42:1466–1470.
Article
12. Chen JR, Lee SY, Yang BH, Lu JJ. Rapid identification and susceptibility testing using the VITEK 2 system using culture fluids from positive BacT/ALERT blood cultures. J Microbiol Immunol Infect. 2008; 41:259–264.
13. Choi J, Jeong HY, Lee GY, Han S, Han S, Jin B, et al. Direct, rapid antimicrobial susceptibility test from positive blood cultures based on microscopic imaging analysis. Sci Rep. 2017; 7:1148.
Article
14. Choi J, Jung YG, Kim J, Kim S, Jung Y, Na H, et al. Rapid antibiotic susceptibility testing by tracking single cell growth in a microfluidic agarose channel system. Lab Chip. 2013; 13:280–287.
Article
15. Choi J, Yoo J, Lee M, Kim EG, Lee JS, Lee S, et al. A rapid antimicrobial susceptibility test based on single-cell morphological analysis. Sci Transl Med. 2014; 6:267ra174.
Article
16. CLSI. Performance standards for antimicrobial susceptibility testing: Twenty-four informational supplement. CLSI document M100-S24. Wayne, PA: Clinical and Laboratory Standards Institute;2014.
17. Mittman SA, Huard RC, Della-Latta P, Whittier S. Comparison of BD Phoenix to Vitek 2, MicroScan MICroSTREP, and Etest for antimicrobial susceptibility testing of Streptococcus pneumoniae. J Clin Microbiol. 2009; 47:3557–3561.
18. Hansen DS, Jensen AG, Nørskov-Lauritsen N, Skov R, Bruun B. Direct identification and susceptibility testing of enteric bacilli from positive blood cultures using VITEK (GNI+/GNS-GA). Clin Microbiol Infect. 2002; 8:38–44.
Article
19. Lupetti A, Barnini S, Morici P, Ghelardi E, Nibbering PH, Campa M. Saponin promotes rapid identification and antimicrobial susceptibility profiling of Gram-positive and Gram-negative bacteria in blood cultures with the Vitek 2 system. Eur J Clin Microbiol Infect Dis. 2013; 32:493–502.
Article
20. Idelevich EA, Schüle I, Grünastel B, Wüllenweber J, Peters G, Becker K. Acceleration of antimicrobial susceptibility testing of positive blood cultures by inoculation of Vitek 2 cards with briefly incubated solid medium cultures. J Clin Microbiol. 2014; 52:4058–4062.
Article
21. Howard WJ, Buschelman BJ, Bale MJ, Pfaller MA, Koontz FP, Jones RN. Vitek GPS card susceptibility testing accuracy using direct inoculation from BACTEC 9240 blood culture bottles. Diagn Microbiol Infect Dis. 1996; 24:109–112.
Article
22. Woods GL, LaTemple D, Cruz C. Evaluation of MicroScan rapid gram-positive panels for detection of oxacillin-resistant staphylococci. J Clin Microbiol. 1994; 32:1058–1059.
Article
Full Text Links
  • ACM
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr