Evaluation of Two Novel Integrated Stand-Alone Spacer Designs Compared with Anterior and Anterior-Posterior Single-Level Lumbar Fusion Techniques: An In Vitro Biomechanical Investigation

Affiliations
  • 1Austin Spine Center, Lakeway Regional Hospital, Lakeway, TX, USA.
  • 2Musculoskeletal Education and Research Center, A Division of Globus Medical, Inc., Audubon, PA, USA. jharris@globusmedical.com

Abstract

STUDY DESIGN: In vitro biomechanical investigation. PURPOSE: To compare the biomechanics of integrated three-screw and four-screw anterior interbody spacer devices and traditional techniques for treatment of degenerative disc disease. OVERVIEW OF LITERATURE: Biomechanical literature describes investigations of operative techniques and integrated devices with four dual-stacked, diverging interbody screws; four alternating, converging screws through a polyether-ether-ketone (PEEK) spacer; and four converging screws threaded within the PEEK spacer. Conflicting reports on the stability of stand-alone devices and the influence of device design on biomechanics warrant investigation.
METHODS
Fourteen cadaveric lumbar spines were divided randomly into two equal groups (n=7). Each spine was tested intact, after discectomy (injured), and with PEEK interbody spacer alone (S), anterior lumbar plate and spacer (AP+S), bilateral pedicle screws and spacer (BPS+S), circumferential fixation with spacer and anterior lumbar plate supplemented with BPS, and three-screw (SA3s) or four-screw (SA4s) integrated spacers. Constructs were tested in flexion-extension (FE), lateral bending (LB), and axial rotation (AR). Researchers performed one-way analysis of variance and independent t-testing (p≤0.05).
RESULTS
Instrumented constructs showed significantly decreased motion compared with intact except the spacer-alone construct in FE and AR (p≤0.05). SA3s showed significantly decreased range of motion (ROM) compared with AP+S in LB (p≤0.05) and comparable ROM in FE and AR. The three-screw design increased stability in FE and LB with no significant differences between integrated spacers or between integrated spacers and BPS+S in all loading modes.
CONCLUSIONS
Integrated spacers provided fixation statistically equivalent to traditional techniques. Comparison of three-screw and four-screw integrated anterior lumbar interbody fusion spacers revealed no significant differences, but the longer, larger-diameter interbody spacer with three-screw design increased stabilization in FE and LB; the diverging four-screw design showed marginal improvement during AR.

Keyword

Intervertebral disc degeneration; Lumbar region; Range of motion; Equipment design

MeSH Terms

Cadaver
Diskectomy
Equipment Design
In Vitro Techniques*
Intervertebral Disc Degeneration
Lumbosacral Region
Pedicle Screws
Range of Motion, Articular
Spine