Korean J Thorac Cardiovasc Surg.  2017 Aug;50(4):291-294. 10.5090/kjtcs.2017.50.4.291.

Delayed Type III Endoleak Caused by Fabric Erosion after Endovascular Repair of an Abdominal Aortic Aneurysm

Affiliations
  • 1Department of Thoracic and Cardiovascular Surgery, Dongguk University Ilsan Hospital, Korea. miracleho@gmail.com

Abstract

A 74-year-old patient presented with recurrent aneurysms in the infrarenal abdominal aorta and right common iliac artery 6 years after endovascular aortic repair using endografts in the same location. The patient underwent an aorto-bi-iliac replacement with removal of the stent graft. Two holes measuring 2 mm each were found in the removed graft, and they appeared to have been caused by wear from continuous friction between the endograft and the aortic wall.

Keyword

Aneurysm; Aorta; Abdominal; Endovascular procedures; Endoleak

MeSH Terms

Aged
Aneurysm
Aorta
Aorta, Abdominal
Aortic Aneurysm, Abdominal*
Blood Vessel Prosthesis
Endoleak*
Endovascular Procedures
Friction
Humans
Iliac Artery
Transplants