Lab Med Online.  2017 Apr;7(2):59-65. 10.3343/lmo.2017.7.2.59.

Analytical Performance of Bio-Rad D-100 on a Hemoglobin A1c Assay

Affiliations
  • 1Department of Laboratory Medicine, Yonsei University College of Medicine, Seoul, Korea. comforter6@yuhs.ac

Abstract

BACKGROUND
Hemoglobin A1c (HbA1c) is considered a marker useful for the follow-up and diagnosis of diabetes and implies the importance of reliable assay methods that are traceable to a reference method. We evaluated analytical performance of a new high-performance liquid chromatography system for the HbA1c assay: D-100 from Bio-Rad Laboratories (USA).
METHODS
We evaluated precision, linearity, and carry-over of D-100, according to the Clinical and Laboratory Standards Institute's guidelines. Comparative analysis of D-100 with Integra 800 (Roche Diagnostics, Germany) and Capillarys 3 (Sebia, France) was conducted. Additionally, we evaluated the throughput of the three instruments.
RESULTS
Precision of low- and high-concentration controls in D-100 showed a CV of less than 1%. The linearity was excellent (R²=0.999) in the range of 3.51-18.7%, and carry-over was not observed. HbA1c results of D-100 (n=144) showed good correlation with those of Integra 800 (r=0.993) and Capillarys 3 (r=0.996). The % bias between D-100 and Integra 800 or Capillarys 3 was within the allowable range at all 3 medical decision levels (5.7%, 6.5%, and 10.0%). Elapsed time in the analysis of the first sample by D-100 was shorter than that of Integra 800 (2.4 vs. 11.1 minutes), but subsequent samples took more time (0.8 vs. 0.3 minutes per sample).
CONCLUSIONS
D-100 showed reliable analytical performance with good precision and linearity, minimal carry-over, and acceptable comparative characteristics relative to other instruments. D-100 is expected to be useful for clinical measurements of HbA1c for diabetes diagnosis and theranostics.

Keyword

Hemoglobin A1c; Bio-Rad D-100; Analytical performance

MeSH Terms

Bias (Epidemiology)
Chromatography, Liquid
Diagnosis
Follow-Up Studies
Methods
Theranostic Nanomedicine

Figure

  • Fig. 1. Linearity of the HbA1c assay using D-100.

  • Fig. 2. A Deming regression plot for D-100 and Integra 800 (A) and for D-100 and Capillarys 3 (B). Abbreviation: CI, confidence interval.

  • Fig. 3. A Bland-Altman difference plot for D-100 and Integra 800 (A) and for D-100 and Capillarys 3 (B).


Reference

1. da Rocha Fernandes J, Ogurtsova K, Linnenkamp U, Guariguata L, Seuring T, Zhang P, et al. IDF Diabetes Atlas estimates of 2014 global health expenditures on diabetes. Diabetes Res Clin Pract. 2016; 117:48–54.
Article
2. Guariguata L, Whiting DR, Hambleton I, Beagley J, Linnenkamp U, Shaw JE. Global estimates of diabetes prevalence for 2013 and projections for 2035. Diabetes Res Clin Pract. 2014; 103:137–49.
Article
3. Bonora E, Tuomilehto J. The pros and cons of diagnosing diabetes with A1C. Diabetes Care. 2011; 34(S2):S184–90.
Article
4. Yun WJ, Shin MH, Kweon SS, Park KS, Lee YH, Nam HS, et al. [A comparison of fasting glucose and HbA1c for the diagnosis of diabetes mellitus among Korean adults]. J Prev Med Public Health. 2010; 43:451–4.
Article
5. Krishnamurti U, Steffes MW. Glycohemoglobin: a primary predictor of the development or reversal of complications of diabetes mellitus. Clin Chem. 2001; 47:1157–65.
Article
6. Gerstein HC, Miller ME, Byington RP, Goff DC Jr, Bigger JT, Buse JB, et al. Effects of intensive glucose lowering in type 2 diabetes. N Engl J Med. 2008; 358:2545–59.
Article
7. Standards of medical care in diabetes–2012. Diabetes Care. 2012; 35(S1):S11–63.
8. Nathan DM. Diabetes: advances in diagnosis and treatment. JAMA. 2015; 314:1052–62.
9. Lee H. The role of HbA1C testing in diagnosing diabetes. Korean J Med. 2010; 79:495–9.
10. Gillett MJ. International Expert Committee report on the role of the A1c assay in the diagnosis of diabetes: Diabetes Care. 2009; 32(7):): 1327-1334. Clin Biochem Rev 2009;30:. 197–200.
11. Little RR. Glycated hemoglobin standardization–National Glycohemoglobin Standardization Program (NGSP) perspective. Clin Chem Lab Med. 2003; 41:1191–8.
Article
12. Weykamp C, John WG, Mosca A, Hoshino T, Little R, Jeppsson JO, et al. The IFCC Reference Measurement System for HbA1c: a 6-year progress report. Clin Chem. 2008; 54:240–8.
Article
13. Sacks DB. Global harmonization of hemoglobin A1c. Clin Chem. 2005; 51:681–3.
Article
14. Braga F, Panteghini M. Standardization and analytical goals for glycated hemoglobin measurement. Clin Chem Lab Med. 2013; 51:1719–26.
Article
15. Diagnosis and classification of diabetes mellitus. Diabetes Care. 2010; 33(S1):S62–9.
16. Standards of medical care in diabetes–2010. Diabetes Care. 2010; 33(S1):S11–61.
17. Icks A, Haastert B, Gandjour A, John J, Lowel H, Holle R, et al. Cost-effectiveness analysis of different screening procedures for type 2 diabetes: the KORA Survey 2000. Diabetes Care. 2004; 27:2120–8.
18. Chung S, Jun SH, Song WH, Song J. Six Years’ Experience of accuracy-based proficiency testing for HbA1c in Korea. J Lab Med Qual Assur. 2015; 37:92–100.
19. Braga F, Dolci A, Montagnana M, Pagani F, Paleari R, Guidi GC, et al. Revaluation of biological variation of glycated hemoglobin (HbA(1c)) using an accurately designed protocol and an assay traceable to the IFCC reference system. Clin Chim Acta. 2011; 412:1412–6.
Article
20. Clinical and Laboratory Standards Institute. Evaluation of precision performance of quantitative measurement methods; approved guideline. 2nd ed.EP5-A2. Wayne, PA: Clinical and Laboratory Standards Institute;2004.
21. Clinical and Laboratory Standards Institute. Evaluation of the linearity of quantitative measurement procedures: a statistical approach; approved guideline. EP6-A. Wayne, PA: Clinical and Laboratory Standards Institute. 2003.
22. Clinical and Laboratory Standards Institute. Method comparison and bias estimation using patient samples; approved guideline. EP9-A2. Wayne, PA: Clinical and Laboratory Standards Institute. 2002.
23. Lee JY, Hong KS, Cho SE. [Comparison of HbA1c analyzers: D-10, Variant II Turbo, Cobas Integra 800, and Afinion AS100]. Korean J Lab Med. 2010; 30:345–50.
Article
24. Clinical and Laboratory Standards Institute. Harmonization of glycohemoglobin measurements; approved guideline. C44-A. Wayne, PA: Clinical and Laboratory Standards Institute. 2002.
25. Sacks DB, Arnold M, Bakris GL, Bruns DE, Horvath AR, Kirkman MS, et al. Executive summary: guidelines and recommendations for laboratory analysis in the diagnosis and management of diabetes mellitus. Clin Chem. 2011; 57:793–8.
Article
26. Jaisson S, Leroy N, Guillard E, Desmons A, Gillery P. Analytical performances of the D-100TM hemoglobin testing system (Bio-Rad) for HbA1c assay. Clin Chem Lab Med. 2015; 53:1473–9.
Article
27. Woodworth A, Korpi-Steiner N, Miller JJ, Rao LV, Yundt-Pacheco J, Kuchi-pudi L, et al. Utilization of assay performance characteristics to estimate hemoglobin A1c result reliability. Clin Chem. 2014; 60:1073–9.
Article
28. Fleming JK. Evaluation of HbA1c on the Roche COBAS Integra 800 closed tube system. Clin Biochem. 2007; 40:822–7.
Article
29. Herpol M, Lanckmans K, Van Neyghem S, Clement P, Crevits S, De Crem K, et al. Evaluation of the Sebia Capillarys 3 Tera and the Bio-Rad D-100 systems for the measurement of hemoglobin A1c. Am J Clin Pathol. 2016; 146:67–77.
Article
30. Rohlfing CL, Parvin CA, Sacks DB, Little RR, Committee NS. Comparing analytic performance criteria: evaluation of HbA1c certification criteria as an example. Clin Chim Acta. 2014; 433:259–63.
Article
31. Little RR. Performance of hemoglobin A1c assay methods: good enough? Clin Chem. 2014; 60:1031–3.
Article
32. Maesa JM, Fernández-Riejos P, Mora CS, de Toro M, Valladares PM, González-Rodriguez C. Evaluation of Bio-Rad D-100 HbA1c analyzer against Tosoh G8 and Menarini HA-8180V. Pract Lab Med. 2016; 5:57–64.
Article
33. Lee K, Kim SM, Jun SH, Song SH, Park KU, Song J. Evaluation of analytical performance of the D-100 hemoglobin testing system for Hemoglobin A1c assay. J Lab Med Qual Assur. 2016; 38:95–101.
Article
Full Text Links
  • LMO
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr