Korean J Thorac Cardiovasc Surg.  2016 Dec;49(6):465-467. 10.5090/kjtcs.2016.49.6.465.

Non-Anastomotic Rupture of a Woven Dacron Graft in the Descending Thoracic Aorta Treated with Endovascular Stent Grafting

Affiliations
  • 1Department of Thoracic and Cardiovascular Surgery, Kyungpook National University School of Medicine, Korea. leejt@knu.ac.kr
  • 2Department of Anesthesiology and Pain Medicine, Kyungpook National University School of Medicine, Korea.

Abstract

The intrinsic structural failure of a Dacron graft resulting from the loss of structural integrity of the graft fabric can cause late graft complications. Late non-anastomotic rupture has traditionally been treated surgically via open thoracotomy. We report a case of the successful use of thoracic endovascular repair to treat a Dacron graft rupture in the descending aorta. The rupture occurred 20 years after the graft had been placed. Two stent grafts were placed at the proximal portion of the surgical graft, covering almost its entire length.

Keyword

Thoracoabdominal aortic aneurysm; Dacron; Rupture; Endovascular procedures

MeSH Terms

Aorta, Thoracic*
Aortic Aneurysm, Thoracic
Blood Vessel Prosthesis*
Endovascular Procedures
Polyethylene Terephthalates*
Rupture*
Stents*
Thoracotomy
Transplants*
Polyethylene Terephthalates
Full Text Links