J Korean Radiol Soc.
1992 Aug;28(4):497-504.
New expandable metallic stents: an experimental study in vessels of dogs
Abstract
- Three-types of expandable metallic stents were constructed to evaluate the differences between the stents : one, formed in a cylindrical zigzag pattern with stainless steel wire (Gianturco stent), another, made by coating the Gianturco stent with silicone rubber(Silicone stent), the third, made by coating the Gianturco stent with 24 carat gold(God stent) for tissue acceptance. A total of 69 stents(each 23 stents of Glanturco, Silicone, and Gold stents) of 8-12 mm in diameter and 20 mm long were placed into normal abdominal aorta, inferior vena cava and iliac artery of nine adult dogs for 2 weeks to 11 months. It was more difficult to introcuce Silicone stent into an introducing sheath than Gianturco or Gold stent due to the thickness of silicone rubber, Four Gianturco stents and three Silicone stents showed migration but Gold stent did not. Luminal narrowing or occlusion was noted in 3 Silicone stents, in 2 Gianturco stents, and in 1 Gold stent. Neointimal proliferation over the stent wires was more rapid and even in God stent than Silicone or Gianturco stent. Although further study is needed, Gold stent seems to be better than Gianturco or Silicone stent as an endovascular graft material.