J Periodontal Implant Sci.  2013 Apr;43(2):101-105.

Comparison of the effect of hand instruments, an ultrasonic scaler, and an erbium-doped yttrium aluminium garnet laser on root surface roughness of teeth with periodontitis: a profilometer study

Affiliations
  • 1Department of Periodontics, Dental School, Shahid Beheshti University of Medical Sciences, Tehran, Iran. mahdi@kadkhodazadeh.ir
  • 2Laser Research Center of Dentistry, Tehran University of Medical Sciences, Tehran, Iran.
  • 3School of Dentistry, Shahid Beheshti University of Medical Sciences, Tehran, Iran.

Abstract

PURPOSE
The present study aimed to measure root surface roughness in teeth with periodontitis by a profilometer following root planning with ultrasonic and hand instruments with and without erbium-doped yttrium aluminium garnet (Er:YAG) laser irradiation.
METHODS
Sixty single-rooted maxillary and mandibular teeth, extracted because of periodontal disease, were collected. The crowns and apices of the roots were cut off using a diamond bur and water coolant. The specimens were mounted in an acrylic resin block such that a plain root surface was accessible. After primary evaluation and setting a baseline, the samples were divided into 4 groups. In group 1, the samples were root planned using a manual curette. The group 2 samples were prepared with an ultrasonic scaler. In group 3, after scaling with hand instrumentation, the roots were treated with a Smart 1240D plus Er:YAG laser and in group 4, the roots were prepared with ultrasonic scaler and subsequently treated with an Er:YAG laser. Root surface roughness was then measured by a profilometer (MahrSurf M300+RD18C system) under controlled laboratory conditions at a temperature of 25degrees C and 41% humidity. The data were analyzed statistically using analysis of variance and a t-test (P<0.05).
RESULTS
Significant differences were detected in terms of surface roughness and surface distortion before and after treatment. The average reduction of the surface roughness after treatment in groups 1, 2, 3, and 4 was 1.89, 1.88, 1.40, and 1.52, respectively. These findings revealed no significant differences among the four groups.
CONCLUSIONS
An Er:YAG laser as an adjunct to traditional scaling and root planning reduces root surface roughness. However, the surface ultrastructure is more irregular than when using conventional methods.

Keyword

Laser therapy; Periodontitis; Tooth; Ultrasonics

MeSH Terms

Crowns
Dental Instruments
Humidity
Laser Therapy
Periodontal Diseases
Periodontitis
Tooth
Ultrasonics
Water
Lasers, Solid-State
Diamond
Water

Reference

1. Walmsley AD, Lea SC, Landini G, Moses AJ. Advances in power driven pocket/root instrumentation. J Clin Periodontol. 2008. 35:8 Suppl. 22–28.
Article
2. Oda S, Nitta H, Setoguchi T, Izumi Y, Ishikawa I. Current concepts and advances in manual and power-driven instrumentation. Periodontol 2000. 2004. 36:45–58.
Article
3. Jacobson L, Blomlof J, Lindskog S. Root surface texture after different scaling modalities. Scand J Dent Res. 1994. 102:156–160.
Article
4. Ribeiro FV, Casarin RC, Nociti FH Junior, Sallum EA, Sallum AW, Casati MZ. Comparative in vitro study of root roughness after instrumentation with ultrasonic and diamond tip sonic scaler. J Appl Oral Sci. 2006. 14:124–129.
Article
5. Casarin RC, Pinto FR, Nociti FH Junior, Sallum EA, Sallum AW, Casati MZ. Assessment of ultrasonic root surface scaling with different power settings: roughness evaluation. Braz J Oral Sci. 2006. 5:996–1000.
6. Maruyama H, Aoki A, Sasaki KM, Takasaki AA, Iwasaki K, Ichinose S, et al. The effect of chemical and/or mechanical conditioning on the Er:YAG laser-treated root cementum: analysis of surface morphology and periodontal ligament fibroblast attachment. Lasers Surg Med. 2008. 40:211–222.
Article
7. de Mendonça AC, Maximo MB, Rodrigues JA, Arrais CA, de Freitas PM, Duarte PM. Er:YAG Laser, ultrasonic system, and curette produce different profiles on dentine root surfaces: an in vitro study. Photomed Laser Surg. 2008. 26:91–97.
Article
8. Trushkowsky RD, Oquendo A. Treatment of dentin hypersensitivity. Dent Clin North Am. 2011. 55:599–608.
Article
9. Sicilia A, Cuesta-Frechoso S, Suarez A, Angulo J, Pordomingo A, De Juan P. Immediate efficacy of diode laser application in the treatment of dentine hypersensitivity in periodontal maintenance patients: a randomized clinical trial. J Clin Periodontol. 2009. 36:650–660.
Article
10. Caruso U, Nastri L, Piccolomini R, d'Ercole S, Mazza C, Guida L. Use of diode laser 980 nm as adjunctive therapy in the treatment of chronic periodontitis: a randomized controlled clinical trial. New Microbiol. 2008. 31:513–518.
11. Ribeiro IW, Sbrana MC, Esper LA, Almeida AL. Evaluation of the effect of the GaAlAs laser on subgingival scaling and root planing. Photomed Laser Surg. 2008. 26:387–391.
Article
12. AboElsaad NS, Soory M, Gadalla LM, Ragab LI, Dunne S, Zalata KR, et al. Effect of soft laser and bioactive glass on bone regeneration in the treatment of infra-bony defects (a clinical study). Lasers Med Sci. 2009. 24:387–395.
Article
13. de Almeida JM, Theodoro LH, Bosco AF, Nagata MJ, Oshiiwa M, Garcia VG. In vivo effect of photodynamic therapy on periodontal bone loss in dental furcations. J Periodontol. 2008. 79:1081–1088.
Article
14. Romanos GE, Henze M, Banihashemi S, Parsanejad HR, Winckler J, Nentwig GH. Removal of epithelium in periodontal pockets following diode (980 nm) laser application in the animal model: an in vitro study. Photomed Laser Surg. 2004. 22:177–183.
Article
15. Cobb CM. Lasers in periodontics: a review of the literature. J Periodontol. 2006. 77:545–564.
Article
16. Kishida M, Sato S, Ito K. Comparison of the effects of various periodontal rotary instruments on surface characteristics of root surface. J Oral Sci. 2004. 46:1–8.
Article
17. Kishida M, Sato S, Ito K. Effects of a new ultrasonic scaler on fibroblast attachment to root surfaces: a scanning electron microscopy analysis. J Periodontal Res. 2004. 39:111–119.
Article
18. Bolortuya G, Ebihara A, Ichinose S, Watanabe S, Anjo T, Kokuzawa C, et al. Effects of dentin surface modifications treated with Er:YAG and Nd:YAG laser irradiation on fibroblast cell adhesion. Photomed Laser Surg. 2012. 30:63–70.
Article
19. Kreisler M, Christoffers AB, Willershausen B, d'Hoedt B. Effect of low-level GaAlAs laser irradiation on the proliferation rate of human periodontal ligament fibroblasts: an in vitro study. J Clin Periodontol. 2003. 30:353–358.
Article
20. Crespi R, Barone A, Covani U, Ciaglia RN, Romanos GE. Effects of CO2 laser treatment on fibroblast attachment to root surfaces. A scanning electron microscopy analysis. J Periodontol. 2002. 73:1308–1312.
Article
21. Miyazaki A, Yamaguchi T, Nishikata J, Okuda K, Suda S, Orima K, et al. Effects of Nd:YAG and CO2 laser treatment and ultrasonic scaling on periodontal pockets of chronic periodontitis patients. J Periodontol. 2003. 74:175–180.
Article
22. Liu CM, Shyu YC, Pei SC, Lan WH, Hou LT. In vitro effect of laser irradiation on cementum-bound endotoxin isolated from periodontally diseased roots. J Periodontol. 2002. 73:1260–1266.
Article
23. Aoki A, Ando Y, Watanabe H, Ishikawa I. In vitro studies on laser scaling of subgingival calculus with an erbium:YAG laser. J Periodontol. 1994. 65:1097–1106.
Article
24. Folwaczny M, Benner KU, Flasskamp B, Mehl A, Hickel R. Effects of 2.94 microm Er:YAG laser radiation on root surfaces treated in situ: a histological study. J Periodontol. 2003. 74:360–365.
Article
25. Noori ZT, Fekrazad R, Eslami B, Etemadi A, Khosravi S, Mir M. Comparing the effects of root surface scaling with ultrasound instruments and Er, Cr:YSGG laser. Lasers Med Sci. 2008. 23:283–287.
Article
26. Rosenberg RM, Ash MM Jr. The effect of root roughness on plaque accumulation and gingival inflammation. J Periodontol. 1974. 45:146–150.
Article
27. Breininger DR, O'Leary TJ, Blumenshine RV. Comparative effectiveness of ultrasonic and hand scaling for the removal of subgingival plaque and calculus. J Periodontol. 1987. 58:9–18.
Article
28. Ritz L, Hefti AF, Rateitschak KH. An in vitro investigation on the loss of root substance in scaling with various instruments. J Clin Periodontol. 1991. 18:643–647.
Article
29. Dragoo MR. A clinical evaluation of hand and ultrasonic instruments on subgingival debridement. 1. With unmodified and modified ultrasonic inserts. Int J Periodontics Restorative Dent. 1992. 12:310–323.
30. Drisko CL. Scaling and root planing without overinstrumentation: hand versus power-driven scalers. Curr Opin Periodontol. 1993. 78–88.
31. Gholami GA, Fekrazad R, Esmaiel-Nejad A, Kalhori KA. An evaluation of the occluding effects of Er; Cr:YSGG, Nd:YAG, CO2 and diode lasers on dentinal tubules: a scanning electron microscope in vitro study. Photomed Laser Surg. 2011. 29:115–121.
Article
Full Text Links
  • JPIS
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr