J Korean Soc Ther Radiol Oncol.
2004 Jun;22(2):155-163.
Tissue Inhomogeneity Correction in Clinical Application of Transmission Dosimetry to Head and Neck Cancer Radiation Treatment
- Affiliations
-
- 1Department of Radiation Oncology, Seoul National University College of Medicine, Korea. snhuh@daum.net
Abstract
- PURPOSE: To confirm the reproducibility of in vivo transmission dosimetry system and the accuracy of the algorithms for the estimation of transmission dose in head and neck radiation therapy patients.
MATERIALS AND MEHTODS: From September 5 to 18, 2001, transmission dose measurements were performed when radiotherapy was given to brain or head and neck cancer patients. The data of 35 patients who were treated more than three times and whose central axis of the beam was not blocked were analyzed in this study. To confirm the reproducibility of this system, transmission dose was measured before daily treatment and then repetitively every hour during the treatment time, with a field size of 10x10 cm2 and a delivery of 100 MU. The accuracy of the transmission dose calculation algorithms was confirmed by comparing estimated dose with measured dose. To accurately estimate transmission dose, tissue inhomogeneity correction was done.
RESULTS
The measurement variations during a day were within +/-0.5% and the daily variations in the checked period were within +/-1.0%, which were acceptable for system reproducibility. The mean errors between estimated and measured doses were within +/-5.0% in patients treated to the brain, +/-2.5% in head, and +/-5.0% in neck.
CONCLUSION
The results of this study confirmed the reproducibility of our system and its usefulness and accuracy for daily treatment. We also found that tissue inhomogeneity correction was necessary for the accurate estimation of transmission dose in patients treated to the head and neck.