Korean J Hepatobiliary Pancreat Surg.  2008 Mar;12(1):41-45.

Bioartificial Liver System using a Fluidized-Bed Bioreactor

Affiliations
  • 1Department of Surgery, Kyungpook National University, College of Medicine, Daegu, Korea. yjhwang@knu.ac.kr
  • 2Department of Neurosurgery, Kyungpook National University, College of Medicine, Daegu, Korea.

Abstract

PURPOSE: Acute fulminant hepatic failure (FHF) is associated with high mortality. Recent studies have shown that a liver support system based on viable hepatocytes can prolong life in animal models of FHF and also in some FHF patients who were successfully bridged to liver transplantation. To be applied in humans, the bioartificial liver (BAL) system should have sufficient liver cell mass to provide adequate bioactive support. The most widely investigated bioreactor at present is based on hollow fiber membrane. However, it is difficult to build a scaledup module of this model. We devised a fluidized-bed bioreactor that is packed with isolated porcine hepatocytes, and these hepatcytes are immobilized in Ca-alginate hydrogel.
METHODS
We isolated fresh porcine hepatocytes using a 2-step collagenase perfusion method, and they were suspended in 1.5% alginate solution. Through a drop-generator, this mixture was gelled in 135mM Cacl2. The resulting spherical beads (mean size: 500 micrometer) were embedded in a module. An average of 2x10(10) hepatocytes were present in the module. The efficacy of our design was tested in pigs that had undergone total hepatic devascularization and portocaval shunt.
RESULTS
The BAL-treated group showed a significantly lower ammonium build-up rate compared to the control group (598.6+/-344.2 microgram/dl vs 1937.6+/-744.1 microgram/dl, respectively, at 8 hours after connecting to BAL). In addition, the intracranial pressure was well controlled in the BAL-treated group, whereas the control group showed a progressive increase of the intracranial pressure (16.9+/-1mmH2O vs 21.9+/-2.6mmH2O, respectively, at 8 hours after connecting to BAL).
CONCLUSION
Our bioartificial liver system is a fluidized-bed bioreactor packed with immobilized porcine hepatocytes, and it seems to be a more effective scaled-up module.

Keyword

Artificial Liver; Acute Liver Failure; Fluidized-bed

MeSH Terms

Alginates
Bioreactors
Collagenases
Glucuronic Acid
Hepatocytes
Hexuronic Acids
Humans
Intracranial Pressure
Liver
Liver Failure, Acute
Liver Transplantation
Liver, Artificial
Membranes
Models, Animal
Perfusion
Quaternary Ammonium Compounds
Swine
Alginates
Collagenases
Glucuronic Acid
Hexuronic Acids
Quaternary Ammonium Compounds