Ventilation Impairment due to Obstruction of a Bifurcated Connector: A case report

Affiliations
  • 1Department of Anesthesiology and Pain Medicine, College of Medicine, Yeungnam University, Daegu, Korea. dhlee415@med.yu.ac.kr

Abstract

Ventilation impairment sometimes developes in patients with a double-lumen tube (DLT) during thoracic anesthesia. A 27-year old man who suffered from a left pneumothorax showed ventilation impairment with right lung obstruction after the induction of anesthesia. We checked breathing sounds before and after clamping each lumen. Correct positioning was confirmed by direct visualization using a flexible fiberoptic bronchoscope and by chest X-ray. The DLT was removed and replaced by a single lumen endotracheal tube because of severe ventilation impairment. On careful examination of the DLT, manufacturing defects in a bifurcated connector were noted, and its openings were occluded with a thin plastic membrane. After replacing the DLT, the ventilation impairment was improved. It is important to test DLT prior to use, and to maintain an awareness that airway obstruction can occur due to structural defects.

Keyword

bifurcated connector; double-lumen tube; obstruction; ventilation impairment

MeSH Terms

Adult
Airway Obstruction
Anesthesia
Bronchoscopes
Constriction
Humans
Lung
Membranes
Plastics
Pneumothorax
Respiratory Sounds
Thorax
Ventilation*
Plastics
Full Text Links
  • KJAE
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr