J Korean Ophthalmol Soc.  2011 Oct;52(10):1161-1166. 10.3341/jkos.2011.52.10.1161.

Comparison of Clinical Results between Ellips and Ozil Modes in Phacoemulsification

Affiliations
  • 1Department of Ophthalmology, College of Medicine, Chungnam National University, Daejeon, Korea. shchoi@cnu.ac.kr
  • 2Chungnam National University Research Institute for Medical Sciences, Daejeon, Korea.

Abstract

PURPOSE
To compare the clinical outcomes following phacoemulsification in Ellips (Signature(R)) and Ozil (Infiniti(R)) modes.
METHODS
Phacoemulsification was performed using Ellips and Ozil modes in 30 eyes each. The phacoemulsification parameters such as the maximum phaco power, vacuum power, and aspiration rates were consistent between the two modes. Nuclear sclerosis grade, average phaco power, and the total phaco energy were recorded. The best-corrected visual acuity, central corneal thickness, and endothelial cell counts were measured preoperatively and on postoperative day 1 and on week 1, 2, 4, and 8.
RESULTS
The average phaco power and total phaco energy were significantly lower in patients with nuclear sclerosis grades 1 and 2 in the Ellips group compared with those in the Ozil group (p < 0.05). However, when these parameters were measured in patients with nuclear sclerosis grades 3 and 4, the difference was not significant between the 2 groups (p > 0.05). Two months postoperatively, the changes in central corneal thickness were 8.73 +/- 17.62 and 5.19 +/- 14.42 microm in the Ellips and Ozil groups, and the changes in endothelial cell count were 100.4 +/- 196.5 and 145.7 +/- 323.8 cells/mm2 (p > 0.05), respectively. The differences did not reach statistical significance.
CONCLUSIONS
Although the ultrasound energy used in the surgical procedure was significantly lower in the Ellips group compared with that in the Ozil group, especially in patients with nuclear sclerosis grades 1 and 2, there were no significant differences in the best-corrected visual acuity, central corneal thickness, and endothelial cell count between the 2 groups.

Keyword

Central corneal thickness; Corneal endothelium; Ellips; Ozil; Transversal phacoemulsification

MeSH Terms

Endothelial Cells
Endothelium, Corneal
Eye
Humans
Phacoemulsification
Sclerosis
Vacuum
Visual Acuity

Cited by  4 articles

Comparison of Clinical Outcomes between High and Low Fluid-Dynamic Parameters during Phacoemulsification
Won Jae Heo, Jin Young Lee, Hong Kyun Kim
J Korean Ophthalmol Soc. 2015;56(12):1860-1866.    doi: 10.3341/jkos.2015.56.12.1860.

Incidence of Posterior Capsular Opacification Based on Low and High Fluid-dynamic Parameters
Yun Hyup Na, Joo Youn Shin, Jong Hyun Lee, Jin Hyoung Kim, Do Hyung Lee
J Korean Ophthalmol Soc. 2016;57(10):1555-1562.    doi: 10.3341/jkos.2016.57.10.1555.

Clinical Effects of an Improved Pump Reaction Rate and Automatic Occlusion Sensing System in Phacoemulsification
You Na Kim, Jin Ah Lee, Jae Yong Kim, Myoung Joon Kim, Hung Won Tchah
J Korean Ophthalmol Soc. 2018;59(11):1017-1023.    doi: 10.3341/jkos.2018.59.11.1017.

Comparison of Clinical Results between 2.2 mm and 2.8 mm Incision Cataract Surgery Using Ellips Ultrasound
Woon Hyung Ghim, Yong Kyun Shin, Kyong Jin Cho
J Korean Ophthalmol Soc. 2015;56(1):19-24.    doi: 10.3341/jkos.2015.56.1.19.


Reference

References

1. Liu Y, Zeng M, Liu X, et al. Torsional mode versus conventional ultrasound mode phacoemulsification: randomized comparative clinical study. J Cataract Refract Surg. 2007; 33:287–92.
2. Schmutz JS, Olson RJ. Thermal comparison of Infiniti OZil and Signature Ellips phacoemulsification systems. Am J Ophthalmol. 2010; 149:762–7.
Article
3. West SK, Rosenthal F, Newland HS, Taylor HR. Use of photographic techniques to grade nuclear cataracts. Invest Ophthalmol Vis Sci. 1988; 29:73–7.
4. Vargas LG, Holzer MP, Solomon KD, et al. Endothelial cell integrity after phacoemulsification with 2 different handpieces. J Cataract Refract Surg. 2004; 30:478–82.
Article
5. O'Brien PD, Fitzpatrick P, Kilmartin DJ, Beatty S. Risk factors for endothelial cell loss after phacoemulsification surgery by a junior resident. J Cataract Refract Surg. 2004; 30:839–43.
6. Polack FM, Sugar A. The phacoemulsification procedure Ⅲ corneal complications. Invest Ophthalmol Vis Sci. 1977; 16:39–46.
7. Gwin RM, Warren JK, Samuelson DA, Gum GG. Effects of phacoemulsification and extracapsular lens removal on corneal thickness and endothelial cell density in the dog. Invest Ophthalmol Vis Sci. 1983; 24:227–36.
8. Walkow T, Anders N, Klebe S. Endothelial cell loss after phacoemulsification: relation to preoperative and intraoperative parameters. J Cataract Refract Surg. 2000; 26:727–32.
Article
9. Bourne RR, Minassian DC, Dart JK, et al. Effect of cataract surgery on the corneal endothelium: modern phacoemulsification compared with extracapsular cataract surgery. Ophthalmology. 2004; 111:679–85.
10. Hayashi K, Hayashi H, Nakao F, Hayashi F. Risk factors for corneal endothelial injury during phacoemulsification. J Cataract Refract Surg. 1996; 22:1079–84.
Article
11. Koch DD, Liu JF, Glasser DB, et al. A comparison of corneal endothelial changes after use of Healon or Viscoat during phacoemulsification. Am J Ophthalmol. 1993; 115:188–201.
Article
12. Kosrirukvongs P, Slade SG, Berkeley RG. Corneal endothelial changes after divide and conquer versus chip and flip phacoemulsification. J Cataract Refract Surg. 1997; 23:1006–12.
Article
13. Gil SY, Kang SB, Lee SH, Chung SK. The effect of phacoemulsification with oscillation device on the cornea and lens opcatiy. J Korean Ophthalmol Soc. 2006; 47:1948–53.
Full Text Links
  • JKOS
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr