J Korean Soc Ther Radiol.
1994 Jun;12(2):219-224.
Total Body Irradiation Technique: Basic Data Measurements and In Vivo Dosimetry
- Affiliations
-
- 1Department of Therapeutic Radiology Catholic University Medical College, Seoul, Korea.
- 2Department of Internal Medicine Catholic University Medical College, Seoul, Korea.
Abstract
-
This paper describes the basic date measurements for total body irradiation with 6 Mv photon beam including compensators designs. The technique uses bilateral opposing field with tissue compensators for the head, neck, lungs, and legs from the hip to toes. In vivo dosimetry was carried out for determining absorbed dose at various regions in 7 patients using diode detectors (MULTIDOSE, Model 9310, MULTIDATA Co., USA). As a results, the dose uniformity of+/-3.5%(generally, within+/-10%) can be achieved with our total body irradiation technique.