J Korean Acad Prosthodont.  2006 Apr;44(2):174-184.

The influence of implant fixture-abutment connection design on screw loosening

Affiliations
  • 1Department of Prosthodontics, Graduate School, Chonnam National University, Korea. psw320@chonnam.ac.kr

Abstract

PURPOSE: Current trend in implant dentistry is changing from external connection to internal connection. To evaluate the splinting of external and internal connection implant on screw loosening, 2-units prosthesis was fabricated with BioPlant System(R) of external connection type and Lifecore STAGE-1 Single Stage Implant System(R) of internal connection type. MATERIAL AND METHOD: Experimental group is classified into three groups. 1) G1-EE: 2-units prosthesis was fabricated with two BioPlant System. of external connection type. 2) G1-EI: 2-units prosthesis was fabricated with one BioPlant System(R) of external connection type and one Lifecore STAGE-1 Single Stage Implant System(R) of internal connection type. 3) G1-II: 2-units prosthesis was fabricated with two Lifecore STAGE-1 Single Stage Implant System(R) of internal connection type. In fabricating 2-units prosthesis, two hexed abutments are recommended when two implants are installed parallel, otherwise one hexed abutment is used on major occlusal force area and one nonhexed abutment is used on the other area. Since it is rare to find two implants being parallel, it is hard to fabricate prosthesis with passive adaptation using two hexed abutments. It is much more difficult to acquire passive adaptation when using hex abutment compared to nonhex abutment. To evaluate the influence of hexed and nonhexed abutment on screw loosening, 2-units prosthesis was fabricated with hexed and nonhexed abutment. Experimental group is classified into three groups. 1) G2-HH: 2-units prosthesis was fabricated with two hexed abutments. 2) G2-HN: 2-units prosthesis was fabricated with one hexed abutment and one nonhexed abutment. 3) G2-NN: 2-units prosthesis was fabricated with two nonhexed abutments. RESULT: The results of comparing the detorque value after loading on a each prosthesis periodically are as follows. 1. In splinting group of external and internal connection implant, G1-II group demonstrated the biggest detorque value, followed by G1-EI group and G1-EE group. 2. There is no notable significance between external connection implant of G1-EI group and G1-EE group and also no significance between internal connection implant of G1-EI group and G1-II group. 3. G2-HH group showed higher detorque value than G2-HN and G2-NN group. From the results, we can concluded that using both external connection and internal connection implant together is clinically acceptable and in order to acquire a good passive adaptation in fabricating 2-units implant prosthesis we can use two nonhexed abutments.

Keyword

Screw loosening; Detorque value; Hexed and nonhexed abutment

MeSH Terms

Bite Force
Dentistry
Prostheses and Implants
Splints
Full Text Links
  • JKAP
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr