J Korean Soc Emerg Med.  2015 Dec;26(6):563-570. 10.0000/jksem.2015.26.6.563.

Comparison of Tracheal Intubation Using the Macintosh Laryngoscope Versus the Disposable Laryngoscope by Inexperienced Personnel: A Manikin Study

Affiliations
  • 1Department of Emergency Medicine, College of Medicine, Soonchunhyang University, Gyeonggido, Korea. emracoon@gmail.com
  • 2Department of Biostatistic Consulting, Soonchunhyang University Medical Center, Gyeonggido, Korea.

Abstract

PURPOSE
Disposable one piece laryngoscope (DOL), which probably carries a lower risk of infection than conventional reusable ones, is the only device which has received approval by the Korean Ministry of Food and Drug safety. In the current study, we evaluated the effectiveness and usefulness of the conventional Macintosh laryngoscope (MAC) and DOL in inexperienced personnel.
METHODS
A randomized crossover trial was designed. Ninety seven participants (Medical students, Paramedic students) were enrolled and performed endotracheal intubation with MAC and DOL in Manikin (SimMan(R) Laedal Medical Corporation, Stravanger; Norway). Cormack-Lehane grade, the number of attempts, time required for intubation, and dental compression were measured to evaluate the effectiveness. The participants were given a questionnaire survey on the usefulness of laryngoscopes (easier to use, lightness, brightness, manageable to use, delivering power, safer for patients, intense to use).
RESULTS
Regarding the aspect of effectiveness, a significant difference in dental compression was observed between DOL and MAC (p=0.011). Others showed no significant differences. Regarding the aspect of usefulness, DOL was superior to MAC except in delivering power.
CONCLUSION
The DOL appears to be a reasonable device for use in emergency airway management. It should be verified for effectiveness and usefulness of newly developed equipment in emergency airway management because it is directly related to life.

Keyword

Disposable equipment; Laryngoscopes; Intubation; Airway management

MeSH Terms

Airway Management
Allied Health Personnel
Disposable Equipment
Emergencies
Humans
Intubation*
Intubation, Intratracheal
Laryngoscopes*
Manikins*
Full Text Links
  • JKSEM
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr