Prog Med Phys.  2016 Mar;27(1):46-53. 10.14316/pmp.2016.27.1.46.

Report of Present Status of Calibration for Domestic Radiation Measurements Instruments

Affiliations
  • 1Department of Radiation Oncology, Kosin University College of Medicine, Busan, Korea.
  • 2Department of Radiation Oncology, Sheikh Khalifa Specialty Hospital Ras Al Khaimah, UAE.
  • 3Department of Radiation Oncology, Gachon University of Medicine and Science, Incheon, Korea.
  • 4Department of Radiation Oncology, Soonchunhyang University College of Medicine, Bucheon, Korea.
  • 5Department of Radiation Oncology, Kwandong University College of Medicine, Seoul, Korea.
  • 6Department of Radiation Oncology, School of Medicine, Ewha Womans University, Seoul, Korea. renalee@ewha.ac.kr
  • 7Department of Radiation Oncology, Yonsei University College of Medicine, Seoul, Korea.

Abstract

Periodical calibrations of radiation detectors are important for accurate quality assurance of therapeutic linac. The measuring instruments such as ion-chamber, thermometer, barometer, and survey meter should be calibrated periodically. Period of calibration for these instruments is suggested 6 month to one year in Korea and two years in other countries nowadays. Therefore, the determination of reasonable period for calibration is needed. In this study, we plan to utilize the results of these survey; frequent in use, how to use and stability of instruments, to determine the optimized period of calibration for the instruments in the departments of radiation oncology in Korea based on the ILAC-G24. The SurveyMonkey web-based survey tool was used and the objects of survey were 18 department of radiation oncology in university hospitals, and 15 departments were answered. The 64 questionnaires which supposed to be answered in 50 minutes were classified as the information of candidates, the thermometer, the barometer, the surveymeter, and the ion-chamber. The thermometers and the barometers were not under periodical calibration for more than half of candidates. The periods of calibration of surveymeters were 6 month to 1 year. We expect that the calibration period can be determined based on these survey results.

Keyword

Calibration period; Uncertainty; Radiation measurement

MeSH Terms

Calibration*
Hospitals, University
Korea
Radiation Oncology
Thermometers
Uncertainty

Figure

  • Fig. 1. 소조사면 측정장비 종류별 분포.


Reference

References

1. Ezzell Gary A., et al. IMRTcommissioning: Multipleinstitution planninganddosimetrycomparisons,areportfromAAPMTask Group119.Med.Phys. 36(11):),. 5359–5373. 2009.
2. Alber M, Broggi S, De Wagter C. Guidelinesforthever-ificationofIMRT. Brussels, Belgium: ESTRO;2008.
3. Thomas J. LoSasso: IMRTDeliverySystemQA. Intensitymodulatedradiationtherapy-Thestateoftheart. Palta JR, Mackie TR, editors. editors).Madison, WI: Med. Phys;p. 561–591. 2003.
4. ICRU 62. Prescribing,Recording,andReportingPhotonBeam Therapy. 2001.
5. Shaw R., Kline M., Gillin L., et al. RadiationTherapy OncologyGroup: radiosurgeryqualityassuranceguidelinesInt JRadiatOncolBiolPhys. 27(5):),. 1231–1239. 1993.
6. NASA Handbook NHB. MetrologyandCa14)libration ProvisionsGuidelines. Jet Propulsion. Laboratory 5300. 4(1A):)(. 2000.
7. Chiaro P.J. Jr. CalibrationIntervalThecnicalBasis Document. InstrumentationandControlsDivision. 1998.
8. ANSI N323A. AmericanNationalStandardRadiation Protection InstrumentationTestandCalibration. PortableSurvey Instruments. IEEE;USA: 1997.
9. IAEA Report NO. 16. CalibrationofRadiationProtection MonitoringInstruments. (. 2000.
10. Notification of Nuclear Safety and Security Commission. Technologicalstandardsforradiationsafetyofmedicalfield (No.2015-005).
11. ILAC-G24. Guidelinesforthedeterminationofcalibrationin-tervalsofmeasuringinstruments. InternationalLaboratory Accreditation Cooperation;2007.
Full Text Links
  • PMP
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr