Improved Activity Estimation using Combined Scatter and Attenuation Correction in SPECT

Abstract

PURPOSE: The pvrpose of this study was to evaluate the accuracy of radioactivity quantitation in Tc-99m SPECT by using combined scatter and attenuation correction.
MATERIALS AND METHODS
A cylindrical phantom which simulates tumors (T) and normal tissue (B) was filled with varying activity ratios of Tc-99m. We acquired emission scans of the phantom using a three-headed SPECT system (Trionix, Inc.) witb two energy windows (photopeak window: 126 154 keV and scatter window: 101 123 keV). We performed the scatter correction with dual-energy window subtraction method (k=0.4) and Chang attenuation correction. Three sets of SPECT images were reconstructed using combined scatter and attenuation correction (SC+AC', attenuation correction (AC) and without any correction (NONE). We compared T/B ratio, irnage contrast [(T-B)/(T+B)) and absolute radioactivity with true values.
RESULTS
SC+AC images had the highest mean values of T/B ratios. Image contrast was 0.92 in SC+AC, which was close to the true value of 1, and higher than AC (0.77) or NONE (0.80). Errors of true activity by SPECT images ranged from 1 to 11% for SC+AC, 22-47% for AC, and 2 16% for NONE in a phantom which was located 2.4cm from the phantom surface. In a phantom located 10,0cm from the surface, SC+AC underestimated by ?4%, NON.E 40%. However, AC overestimated by 10%.
CONCLUSION
We conclude that accurate SPECT activity quantitation of Tc-99m distribution can be achieved by dual window scatter correc.tion combined with attenuation correction.

Keyword

SPECT; Radioactivity; Scatter correction; Attenuation correction