A Study on the Visual System of the Digital Periapical Images
- Affiliations
-
- 1Department of Oral and Maxillofacial Radiology, School of Dentistry, Chonbuk Natiional University, Korea.
- 2Institute of Oral Bio Science Chonbuk Natiional University, Korea.
Abstract
OBJECTIVES
The purpose of this study was to evaluate the optimal distance and angle of observers by modulation transfer functions(MTFs) and receiver operating characteristics(ROCs).
MATERIALS AND METHODS
Digital periapical radiograms were taken from 43 patients who have dental diseases(19 patients : dental caries, 12 patients : periapical lesions, 12 patients : periodontal diseases). Segmental images(4x4cm) were evaluated by 4 MTFs and ROC analysis.
RESULTS
The optimal distance(magnification) using MTF by Mannos & Sakrison was 12.97, and those by Nill, Ngan and Rao were 8.39, 4.78, 5.84 respectively. The optimal distance obtained from 4 radiologists by ROC analysis was 32cm(Az value : 0.89), and it was 40cm(Az value : 0.78) from 4 non-radiologists. There were significant differences of Az values between 4 radiologists and 4 non-radiologists at 24, 32 and 40cm (p<0.05). No significant differences of optimal distances were observed using 4 MTFs among +20, +10, 0, -10, -20 degrees(p>0.05). The optimal angle obtained from 4 radiologists by ROC analysis was +20 degrees (Az value : 0.91), and no significant differences of Az values were observed among +20, +10, 0, -10 and 20 degrees(p>0.05). The optimal angle obtained from 4 non-radiologists by ROC analysis was 0 degree(Az value : 0.81), and no significant differences of Az values were observed among +20, +10, 0, -10 and 20 degrees(p>0.05). And there was significant difference of Az value between 4 radiologists and 4 non-radiologists at +20 degree(p<0.05), but no significant differences of Az values were observed among +10, 0, -10 and -20 degrees(p>0.05).