The Study on the Use of a Cylindrical Ionization Chamber for the Calibration of a 6 MeV Electron Beam

Affiliations
  • 1Department of Radiation Oncology, College of Medicine, Hanyang University, Seoul, Korea.
  • 2Department of Radiation Oncology, College of Medicine, Inha University, Korea.
  • 3Department of Radiation Oncology, Gachon University Gil Hospital, Incheon, Korea.
  • 4National Institute of Food and Drug Safety Evaluation, Korea.
  • 5Department of Radiation Oncology, College of Medicine, Kyung Hee University, Seoul, Korea. ohsd32@naver.com

Abstract

The standard dosimetry systems based on an absorbed dose to water recommend to use a planeparallel chamber for the calibration of such a low-megavoltage electron beam as a nominal energy of 6 MeV. For this energy ranges of an electron beam a cylindrical chamber should not be used for the routinely regular beam calibration, but the feasibility of the temporary use of a cylindrical chamber was studied to give temporary solutions for special situations users meet. The PTW30013 chambers and the electron beam quality of R(50)=2.25 g/cm2 were selected for this study. 10 PTW30013 chambers, a cylindrical type of chamber, were calibrated in KFDA, the secondary standards dosimetry laboratories, and given the absorbed dose-to-water calibration factors, respectively. A "temporary" kappa(Q,Q0) for each chamber were calculated using the absorbed dose determined by a cross-calibrated planeparallel chamber, with the result of an average 0.9352 for 10 chambers. This value for PTW30013 chamber was used to determine an absorbed dose to water at the reference depth. The absorbed doses determined by PTW30013 chambers were in an agreement within 2% with that by ROOS chamber. In a certain situation where a cylindrical chamber be used instead of a planeparellel chamber, the value of 0.9352 might be useful to determine an absorbed dose to water in the same beam quality of electron beam as this study.

Keyword

Absorbed dose to water calibration factor; Beam quality correction factor; Cylindrical chamber; Low-megavoltage electron beam

MeSH Terms

Calibration
Electrons
Water
Water
Full Text Links
  • KJMP
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr