Korean J Med Phys.
2009 Jun;20(2):97-105.
Study on Absorbed Dose Determination of Electron Beam Quality for Cross-calibration with Plane-parallel Ionization Chamber
- Affiliations
-
- 1Proton Therapy Center, National Cancer Center, Korea. cool_park@ncc.re.kr
- 2Department of Radiation Oncology, School of Medicine, Kyung Hee University, Seoul, Korea.
- 3Department of Biomedical Engineering, The Catholic University of Korea, Seoul, Korea.
Abstract
- Absorbed dose to water based protocols recommended that plane-parallel chambers be calibrated against calibrated cylindrical chambers in a high energy electron beam with R50>7 g/cm2 (E> or =16 MeV). However, such high-energy electron beams are not available at all radiotherapy centers. In this study, we are compared the absorbed dose to water determined according to cross-calibration method in a high energy electron beam of 16 MeV and in electron beam energies of 12 MeV below the cross-calibration quality remark. Absorbed dose were performed for PTW 30013, Wellhofer FC65G Farmer type cylindrical chamber and for PTW 34001, Wellhofer PPC40 Roos type plane-parallel chamber. The cylindrical and the plane-parallel chamber to be calibrated are compared by alternately positioning each at reference depth, zref=0.6R50-0.1 in water phantom. The DW of plane-parallel chamber are derived using across-calibration method at high-energy electron beams of 16, 20 MeV. Then a good agreement is obtained the DW of plane-parallel chamber in 12 MeV. The agreement between 20 MeV and 12 MeV are within 0.2% for IAEA TRS-398.