Imaging Sci Dent.  2013 Jun;43(2):63-69. 10.5624/isd.2013.43.2.63.

Patient radiation dose and protection from cone-beam computed tomography

Affiliations
  • 1Department of Oral and Maxillofacial Radiology, Peking University School and Hospital of Stomatology, Beijing, China. kqgang@bjmu.edu.cn

Abstract

After over one decade development, cone beam computed tomography (CBCT) has been widely accepted for clinical application in almost every field of dentistry. Meanwhile, the radiation dose of CBCT to patient has also caused broad concern. According to the literature, the effective radiation doses of CBCTs in nowadays market fall into a considerably wide range that is from 19 microSv to 1073 microSv and closely related to the imaging detector, field of view, and voxel sizes used for scanning. To deeply understand the potential risk from CBCT, this report also reviewed the effective doses from literatures on intra-oral radiograph, panoramic radiograph, lateral and posteroanterior cephalometric radiograph, multi-slice CT, and so on. The protection effect of thyroid collar and leaded glasses were also reviewed.

Keyword

Radiation Dosage; Cone-Beam Computed Tomography; Tomography, Spiral Computed; Radiation Protection